FDA Enforcement Class II Ongoing

Omega Medical Imaging Elevating Monitor Suspension, Part Number 1000-0085

Recall: Z-0156-2024 · Reported November 1, 2023

Enforcement

Recall Number
Z-0156-2024
Event ID
93018
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Omega Medical Imaging, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
November 1, 2023
Initiation Date
February 28, 2023
Classification Date
October 24, 2023
Address
3400 Saint Johns Pkwy Ste 1020, N/A, Sanford, FL, 32771-6769, United States

Description

Omega Medical Imaging Elevating Monitor Suspension, Part Number 1000-0085

Reason

An actuator separated from the pivot mechanism on a lower monitor boom system column and became detached and was restrained by the installed safety cable at one location.

Code Info

all units manufactured between 10-22-2012 and 07-09-2019

Distribution

Worldwide distribution - US Nationwide and the country of Pakistan.

Quantity

52 units