FDA Enforcement
Class II
Terminated
V-Twin, Catalog # 6002-800, UDI # 03661540600180 Product Usage: automated chemistry analyzer intended for clinical use
Recall: Z-0156-2020
·
Reported October 23, 2019
Enforcement
- Recall Number
- Z-0156-2020
- Event ID
- 83582
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Vital Scientific N.V.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- October 23, 2019
- Initiation Date
- January 10, 2012
- Classification Date
- October 16, 2019
- Termination Date
- April 19, 2024
- Address
- 24, Kanaalweg, Postbus 100, Dieren, N/A, N/A, Netherlands
Description
V-Twin, Catalog # 6002-800, UDI # 03661540600180 Product Usage: automated chemistry analyzer intended for clinical use
Reason
The action was initiated due power supply failure. During a power supply failure the instruments malfunctions and no patient results are produced.
Code Info
serial numbers: 10-3331; 10-3332; 11-3301; 11-3304 through 11-3306; 11-3308 through 11-3312; 11-3315; 11-3316; 11-3318 through 11-3342; 12-3302.
Distribution
US Nationwide in the states of NY, and UT
Quantity
39 instruments