FDA Enforcement Class II Terminated

V-Twin, Catalog # 6002-800, UDI # 03661540600180 Product Usage: automated chemistry analyzer intended for clinical use

Recall: Z-0156-2020 · Reported October 23, 2019

Enforcement

Recall Number
Z-0156-2020
Event ID
83582
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Vital Scientific N.V.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
October 23, 2019
Initiation Date
January 10, 2012
Classification Date
October 16, 2019
Termination Date
April 19, 2024
Address
24, Kanaalweg, Postbus 100, Dieren, N/A, N/A, Netherlands

Description

V-Twin, Catalog # 6002-800, UDI # 03661540600180 Product Usage: automated chemistry analyzer intended for clinical use

Reason

The action was initiated due power supply failure. During a power supply failure the instruments malfunctions and no patient results are produced.

Code Info

serial numbers: 10-3331; 10-3332; 11-3301; 11-3304 through 11-3306; 11-3308 through 11-3312; 11-3315; 11-3316; 11-3318 through 11-3342; 12-3302.

Distribution

US Nationwide in the states of NY, and UT

Quantity

39 instruments