FDA Enforcement
Class II
Terminated
OneTray Sealed Sterilization Containers Product Usage: A rigid sterilization container used to steam sterilize medical devices
Recall: Z-0153-2016
·
Reported October 28, 2015
Enforcement
- Recall Number
- Z-0153-2016
- Event ID
- 72011
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Innovative Sterlization Technologies LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Visit
- Report Date
- October 28, 2015
- Initiation Date
- February 3, 2014
- Classification Date
- October 21, 2015
- Termination Date
- February 4, 2016
- Address
- 7625 Paragon Rd Ste A, N/A, Dayton, OH, 45459-4063, United States
Description
OneTray Sealed Sterilization Containers Product Usage: A rigid sterilization container used to steam sterilize medical devices
Reason
The wrong sterilization time was on the label. The label lists Steam Gravity 10 - 34 minutes at 270 degrees F instead of Steam Gravity 17 - 34 minutes at 270 degrees F.
Code Info
M2104 Lot # 044305, 044379, 044380 & 044418. M2408 Lot # 044296, 022304, 044381, 044420, 044421 & 044422.
Distribution
US Nationwide Distribution in the states of IL, IN, LA, OH, PA, TX & WV.
Quantity
M2104 - 25 units; M2408 - 84 units