Freedom EVO-2 100 Base Unit, with Air LiHa Arms The Freedom EVO is an open automation platform product for general laboratory use. It is intended for routine laboratory tasks, such as general purpose pipetting and general purpose liquid handling and robotic processes.

Recall: Z-0149-2015 · Reported November 5, 2014

Enforcement

Recall Number
Z-0149-2015
Event ID
69369
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Tecan US, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Telephone
Report Date
November 5, 2014
Initiation Date
September 26, 2014
Classification Date
October 29, 2014
Termination Date
July 10, 2017
Address
9401 Globe Center Drive, Suite 140, Morrisville, NC, 27560, United States

Description

Freedom EVO-2 100 Base Unit, with Air LiHa Arms The Freedom EVO is an open automation platform product for general laboratory use. It is intended for routine laboratory tasks, such as general purpose pipetting and general purpose liquid handling and robotic processes.

Reason

Distortion (noise interference) of the liquid level capacitance can result in air aspiration instead of liquid aspiration. This might cause a false signal (air detection) with possible dispense errors resulting in erroneous results.

Code Info

Part Number: 10641100, SN: 1407006311.

Distribution

US Distribution to IL.

Quantity

2 units