siemens Dimension¿ Phenytoin Flex¿ Reagent Cartridge The PTN method used on the Dimension¿ clinical chemistry system is an in vitro diagnostic test intended to measure phenytoin (dilantin, diphenylhydantoin), an anti-epileptic drug, in human serum and plasma. PTN test results are used in the diagnosis and treatment of phenytoin overdose and in monitoring levels of phenytoin to ensure appropriate therapy.

Recall: Z-0148-2015 · Reported November 5, 2014

Enforcement

Recall Number
Z-0148-2015
Event ID
69372
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Siemens Healthcare Diagnostics, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 5, 2014
Initiation Date
September 24, 2014
Classification Date
October 29, 2014
Termination Date
August 17, 2016
Address
500 Gbc Dr Ms 514, PO BOX 6101, Newark, DE, 19702-2466, United States

Description

siemens Dimension¿ Phenytoin Flex¿ Reagent Cartridge The PTN method used on the Dimension¿ clinical chemistry system is an in vitro diagnostic test intended to measure phenytoin (dilantin, diphenylhydantoin), an anti-epileptic drug, in human serum and plasma. PTN test results are used in the diagnosis and treatment of phenytoin overdose and in monitoring levels of phenytoin to ensure appropriate therapy.

Reason

Internal testing has confirmed a low absorbance range between the levels 1 and 5 calibrators with this lot resulting in the potential for a positive or negative bias and imprecision across the PTN assay range.

Code Info

lot FA5154, exp. 2015-06-03

Distribution

Worldwide Distribution - USA (nationwide) including Puerto Rico., and the countries of Australia, Brazil, Bahamas, Canada, Chile, India, Japan and Uruguay

Quantity

1733