FDA Enforcement Class I Ongoing

Hillrom Hospital Bed Accessory, WatchCare Incontinence Management System for VersaCare Bed Rev. K, Product Code/Part Number P00697902

Recall: Z-0144-2023 · Reported November 9, 2022

Enforcement

Recall Number
Z-0144-2023
Event ID
90865
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
Baxter Healthcare Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 9, 2022
Initiation Date
September 30, 2022
Classification Date
October 31, 2022
Address
1 Baxter Pkwy, N/A, Deerfield, IL, 60015-4625, United States

Description

Hillrom Hospital Bed Accessory, WatchCare Incontinence Management System for VersaCare Bed Rev. K, Product Code/Part Number P00697902

Reason

Radio frequency emission from a functioning WatchCare device may potentially impact other devices (including, but not limited to telemetry devices, bladder scanner, fetal monitor/doppler, infusion pumps, and insulin pumps).

Code Info

UDI/DI 887761998872, all serial numbers

Distribution

US

Quantity

63 units