FDA Enforcement
Class I
Ongoing
Hillrom Hospital Bed Accessory, Watch Care Incontinence Management System for Centrella Bed, Product Code/Part Number P00697905
Recall: Z-0142-2023
·
Reported November 9, 2022
Enforcement
- Recall Number
- Z-0142-2023
- Event ID
- 90865
- Classification
- Class I
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Baxter Healthcare Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- November 9, 2022
- Initiation Date
- September 30, 2022
- Classification Date
- October 31, 2022
- Address
- 1 Baxter Pkwy, N/A, Deerfield, IL, 60015-4625, United States
Description
Hillrom Hospital Bed Accessory, Watch Care Incontinence Management System for Centrella Bed, Product Code/Part Number P00697905
Reason
Radio frequency emission from a functioning WatchCare device may potentially impact other devices (including, but not limited to telemetry devices, bladder scanner, fetal monitor/doppler, infusion pumps, and insulin pumps).
Code Info
UDI/DI 887761984998, all serial numbers
Distribution
US
Quantity
814 units