FDA Enforcement Class II Terminated

Canon PET-CT SCANNER, Cartesion Prime, MODEL PCD-1000A

Recall: Z-0142-2022 · Reported October 27, 2021

Enforcement

Recall Number
Z-0142-2022
Event ID
88691
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Canon Medical System, USA, INC.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
October 27, 2021
Initiation Date
November 11, 2020
Classification Date
October 19, 2021
Termination Date
June 30, 2023
Address
2441 Michelle Dr, N/A, Tustin, CA, 92780-7047, United States

Description

Canon PET-CT SCANNER, Cartesion Prime, MODEL PCD-1000A

Reason

A software problem has been identified which could result in the diagnostic imaging system not proceeding to the next actual scan even though automatic start of the next scan is specified. This could result in the diagnostic imaging system failing resulting in rescanning and reinjection of contrast medium.

Code Info

Affected serial numbers: 3DA1982001, 3DA2072002, 3DA2072003

Distribution

US Nationwide distribution in the states of NV, AR MN.

Quantity

3 units