Trinias unity, Digital Angiography System with Catheterization Table, Model: KS-100 This system consists of the following main components: C arm, catherization table, image processing equipment, x-ray high voltage generator, x-ray tube. There are several types of each component.
Enforcement
- Recall Number
- Z-0139-2025
- Event ID
- 95406
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 30, 2024
- Initiation Date
- September 6, 2024
- Classification Date
- October 22, 2024
- Address
- Nishinokyo Kuwabara -Cho, Nakagyo-Ku; 1, Kyoto, N/A, N/A, Japan
Description
Trinias unity, Digital Angiography System with Catheterization Table, Model: KS-100 This system consists of the following main components: C arm, catherization table, image processing equipment, x-ray high voltage generator, x-ray tube. There are several types of each component.
If digital angiography system is used with a specific catheterization table and the operator directs simultaneous movement of the table and the C-arm using direct memory, then direct memory button is released, the table may continue to move, which the operator can stop by pressing the stop button.
UDI-DI: 04540217058327, 04540217049066, 04540217058143 Serial Numbers: 41EFCD7AC001, 41E58C773001, 41EFCD79C001, 41EFCD78B001, 41EFC73A9001, 41EFCD7D4001, 41E58C76B001, 41EFC73D9001, 41EFCD77C001, 41EFC73A5001, 41EFC73C5001, 41EFCD79A001, 41EFCD7A6001.
US Nationwide distribution in the states of MA, IL, SC, KY, SC, OH, TN.