FDA Enforcement Class II Terminated

6.5mm Cancellous Bone Screw Intended to be used for the fixation of the acetabular shell or fracture repair.

Recall: Z-0139-2013 · Reported October 31, 2012

Enforcement

Recall Number
Z-0139-2013
Event ID
63321
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Encore Medical, Lp
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 31, 2012
Initiation Date
September 28, 2012
Classification Date
October 25, 2012
Termination Date
January 25, 2013
Address
9800 Metric Blvd, N/A, Austin, TX, 78758, United States

Description

6.5mm Cancellous Bone Screw Intended to be used for the fixation of the acetabular shell or fracture repair.

Reason

A product complaint was received where a size 20 screw was labeled as size 30. After receipt, measurement of the screw determined it to be a size 25. Further investigation also determined that there were other screws labeled as size 25, but a size 30 screw was inside of the package. It was thought that all of these screws were found in-house and were contained, but it was determined later that on

Code Info

Model Number 10-55-025, Lot Number 006A1057.

Distribution

Nationwide Distribution only to CA.

Quantity

1 unit