FDA Enforcement Class II Terminated

CORE¿ Trumpet Handpiece only, Catalog Number CD8450 The CORE Trumpet (CD81XX, CD83XX, and CD84XX) has applications in gynecological laparoscopy, laparoscopic cholecystectomy and other laparoscopic procedures

Recall: Z-0135-2017 · Reported October 26, 2016

Enforcement

Recall Number
Z-0135-2017
Event ID
74774
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
ConMed Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 26, 2016
Initiation Date
June 27, 2016
Classification Date
October 18, 2016
Termination Date
June 7, 2018
Address
525 French Rd, N/A, Utica, NY, 13502-5945, United States

Description

CORE¿ Trumpet Handpiece only, Catalog Number CD8450 The CORE Trumpet (CD81XX, CD83XX, and CD84XX) has applications in gynecological laparoscopy, laparoscopic cholecystectomy and other laparoscopic procedures

Reason

The packaging seal may contain a crease. The crease may result in an open channel. These channels may compromise sterility of the product.

Code Info

Beginning Lot Code 110609X to Ending Lot Code 20160308X

Distribution

Worldwide Distribution - US (Nationwide) Internationally: to Australia, Austria, Uruguay, Denmark, Germany, Ireland, Italy, United Kingdom, Guatemala, Czech Republic & Canada.

Quantity