da Vinci X (IS4200) and Xi (IS4000) systems, which consists of finished devices Surgeon Console (SSC), Patient Cart (PSCART), Vision System (VSS)
Enforcement
- Recall Number
- Z-0133-2023
- Event ID
- 90826
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Intuitive Surgical, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- November 2, 2022
- Initiation Date
- September 20, 2022
- Classification Date
- October 25, 2022
- Address
- 1266 Kifer Rd Bldg 101, N/A, Sunnyvale, CA, 94086-5304, United States
Description
da Vinci X (IS4200) and Xi (IS4000) systems, which consists of finished devices Surgeon Console (SSC), Patient Cart (PSCART), Vision System (VSS)
Inadvertent energy delivery from surgical system instrument if 1) Force bipolar and bipolar instruments installed on system 2) Force bipolar connected to Force Triad generator, bipolar not connected to generator 3) Yellow pedal associated with force bipolar is pressed, released with head in surgeon console 4) Blue pedal associated with bipolar is pressed, resulting in force bipolar energy delivery
Software version P10 (OS4 v10.0.0/A70_P10_B738) SOFTWARE,EMBEDDED RLS,IS4000 AND IS4200,A70_P10_B738 (Model: 610092-738) on Model Name/Model Number/UDI-DI: ASSY,PSCART,IS4000,4-ARM,P10/380652-55/00886874110720, ASSY,SSC,IS4000,P10/380677-28/00886874110744, ASSY,VSS VISION SYSTEM,IS4000,P10/381121-45/00886874110898, ASSY,PSCART,IS4200,P10/380620-47/00886874115404, ASSY,VSC,IS4000,RECONDITIONED,P10/380721-15/00886874115343, ASSY,SSC,IS4000,RECONDITIONED,P10/380723-16/00886874115374
US: MS, PA, WI, KY, TN, MI, FL, GA, NC, NV, LA, OH, IL, MO, IN, CA, AK, MD, TX, IA, MA, NE, AZ, WA, CO, AL, SC
162