The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field, to magnify the visual field using the magnifying feature of the iPhone, while simultaneously selecting a single number on the keyboard.
Enforcement
- Recall Number
- Z-0132-2026
- Event ID
- 97597
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Insulet Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- October 22, 2025
- Initiation Date
- September 11, 2025
- Classification Date
- October 10, 2025
- Address
- 50 & 100 Nagog Park, Acton, MA, 01720-3440, United States
Description
The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field, to magnify the visual field using the magnifying feature of the iPhone, while simultaneously selecting a single number on the keyboard.
The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field, to magnify the visual field using the magnifying feature of the iPhone, while simultaneously selecting a single number on the keyboard. This causes the single digit number to duplicate. This could result in errors in manual entry of digits - examples: bolus entry, I:C ratio, bolus calculator, prior to starting the bolus insulin delivery. Therefore, this may result in over delivery of insulin which may cause hypoglycemia.
Omnipod 5 App Software. UDI: 10385083000459. All software versions 2.X.X
US Nationwide distribution.
174,013 units