FDA Enforcement Class II Terminated

Dimension Vista¿ HbA1c

Recall: Z-0127-2015 · Reported November 5, 2014

Enforcement

Recall Number
Z-0127-2015
Event ID
69334
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Siemens Healthcare Diagnostics, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 5, 2014
Initiation Date
August 14, 2014
Classification Date
October 24, 2014
Termination Date
August 10, 2015
Address
500 Gbc Dr Ms 514, PO BOX 6101, Newark, DE, 19702-2466, United States

Description

Dimension Vista¿ HbA1c

Reason

In March, Siemens Healthcare Diagnostics issued an Urgent Medical Device Correction for decreased recovery of patient samples with the Dimension Vista® CSA & CSAE Flex® reagent cartridges when whole blood samples are run from Small Sample Containers (SSC). Siemens has confirmed that when whole blood samples are run from SSC, HbA1c values may exhibit a percent bias ranging from - 8.2% t

Code Info

catalog #K3105A, material #10470481, All in-date lots

Distribution

Worldwide Distribution: US (nationwide) and country of: Canada.

Quantity

20043