FDA Enforcement
Class II
Terminated
Dimension Vista¿ HbA1c
Recall: Z-0127-2015
·
Reported November 5, 2014
Enforcement
- Recall Number
- Z-0127-2015
- Event ID
- 69334
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Siemens Healthcare Diagnostics, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- November 5, 2014
- Initiation Date
- August 14, 2014
- Classification Date
- October 24, 2014
- Termination Date
- August 10, 2015
- Address
- 500 Gbc Dr Ms 514, PO BOX 6101, Newark, DE, 19702-2466, United States
Description
Dimension Vista¿ HbA1c
Reason
In March, Siemens Healthcare Diagnostics issued an Urgent Medical Device Correction for decreased recovery of patient samples with the Dimension Vista® CSA & CSAE Flex® reagent cartridges when whole blood samples are run from Small Sample Containers (SSC). Siemens has confirmed that when whole blood samples are run from SSC, HbA1c values may exhibit a percent bias ranging from - 8.2% t
Code Info
catalog #K3105A, material #10470481, All in-date lots
Distribution
Worldwide Distribution: US (nationwide) and country of: Canada.
Quantity
20043