FDA Enforcement Class III Terminated

MR Coils The MR Coil is intended to be used in conjunction with a Magnetic Resonance Scanner to produce diagnostic images of the anatomy of interest that can be interpreted by a trained physician.

Recall: Z-0126-2021 · Reported October 28, 2020

Enforcement

Recall Number
Z-0126-2021
Event ID
86501
Classification
Class III
Status
Terminated
Product Type
Devices
Firm
Invivo Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 28, 2020
Initiation Date
September 14, 2020
Classification Date
October 19, 2020
Termination Date
August 14, 2023
Address
3545 SW 47th Ave, N/A, Gainesville, FL, 32608-7691, United States

Description

MR Coils The MR Coil is intended to be used in conjunction with a Magnetic Resonance Scanner to produce diagnostic images of the anatomy of interest that can be interpreted by a trained physician.

Reason

The labels have been mistakenly printed with an IPX1 rating symbol. IPX ratings relate to protections that products have from water ingress.

Code Info

Models: dS Knee Sch 3.0T, dS Knee Sch 1.5T, dS HiRes Hand/Wrist 16ch 1.5T, dS HiRes Hand/Wrist, dS Shoulder 16ch 3, dS Shoulder 16ch 1.5T, dS Shoulder Sch 1.ST,dS Shoulder Sch 3.0T. Part Numbers: 4522-150-40202, 4522-150-40362, 4535-300-54581, 4535-300-71071, 4535-300-71271, 4535-300-71281, 535-302-77861, 4535-302-77871, 4535-302-77891, 4535-302-94611, 4535-302-94612, 4535-302-94613, 4535-302-94621, 4535-302-94623, 4535-303-59931, 4535-303-59951, 4535-303-59953, 4535-303-59971, 4535-303-59973, 4535-303-60031, 4535-303-60033, 4535-303-60051, 4535-303-60053, 4535-303-60071, 4535-303-60073, 4535-303-60751, 4535-303-60771, 4598-000-97974, 4598-000-97984, 4598-000-97994, 4598-012-90041, 4598-012-90042, 4598-012-90051, 598-012-90061, 4598-012-95531, 4598-013-12411, 4598-014-99951, 4598-014-99971, 4598-014-99991, 4598-015-00031, 4598-015-00051, 4598-015-00071, 9896-030-13432, 9896-030-13972, 9896-030-24282, 9896-030-24402, 4598-012-91131, 4598-012-95481, 9896-032-00441, 9896-032-00442, 9896-032-00443, 9896-032-00451, 9896-032-00453, 9896-032-08011, 9896-032-08021, 9896-032-08031, 9896-032-08051, 9896-032-08061, 9896-032-08071, 9896-032-15621, 9896-032-15671, 9896-032-15681, 9896-032-15682, 9896-032-15691, 9896-032-15692, 9896-032-15731, 9896-032-15741, 9896-032-15751, 4598-012-90062, 4598-012-90091, 4598-012-90101, 4598-012-90102, 4598-012-90121, 4598-012-90122, 4598-012-90261, 4598-012-90601, 4598-012-90631, 4598-012-90831, 4598-012-90891, 4598-012-90892, 9896-032-15821, 9896-032-15822, 9896-032-16021, 9896-032-16031, 9896-032-16032, 9896-032-16041, 9896-032-16042, 9896-032-16091, 9896-032-16161, 9896-032-16301, 9896-032-16321. Serial Numbers: 5250, 5468, 50474, 5275, 5280, 5244, 5282, 5497,5253, 5503,5255, 50512, 5516, 5518, 5256, 50510, 5509, 5278, 5508, 50525, 5361, 5532, 50545, 50543, 5272, 2020, 5530, 50549, 5311, 2019, 5309, 50555, 50552, 5786, 5438, 5322, 50589, 5591, 5447, 5587, 5327, 50599, 5597, 50591, 5598, 5341, 5809, 5413, 5463, 50612, 5629, 5340, 5416, 5616, 50626, 5345, 5348, 5648, 5471, 5360, 5484, 5816, 5429, 5798, 5497, 5426, 5675, 5367, 50671, 5640, 5352, 50658, 5693, 5481, 5828, 5446, 5641, 5350, 50667, 5682, 5643, 50655, 50650, 5686, 5380, 5639, 50653, 50694, 5719, 5477, 50761, 5751, 5497, 50721, 5499, 5757, 5755, 50772, 5509, 50750, 5772, 5511, 5574, 5370, 50797, 17888, 5839, 5503, 5589, 5746, 50787, 50785, 17898, 5502, 5579, 5877, 5876, 5786, 5702, 5469, 50821, 5781, 5523, 5813, 5897, 50846, 17943, 5873, 5558, 5818, 17958, 5502, 50873, 17969, 50864, 17952, 50868, 17956, 5536, 5490, 18011, 5896, 5850, 5904, 5849, 5616, 5899, 5621, 5884, 5857, 5615, 5891, 5833, 5614, 5887, 5599, 5585, 5943, 5891, 17983, 5578, 5902, 5875, 5638, 18034, 50956, 5924, 5878, 5632, 50986, 18081, 5540, 50959, 18095, 50962, 5651, 5911, 5956, 5554, 5555, 5556, 18010, 18057, 50980, 1, 14, 10, 10, 5933, 5903, 5648, 5547, 50967, 5568, 18046, 18061, 50997, 17989, 51024, 5563, 5567, 5561, 18072, 19, 13, 14, 18071, 51022, 18068, 51009, 5566, 51030, 5675, 5565, 18160, 51047, 5596, 18096, 51002, 51004, 18138, 18105, 5601, 51006, 18161, 18168, 5629, 18106, 5632, 18112, 5628, 5572, 18080, 11, 11, 9, 9, 5593, 18131, 51059, 51057, 5591, 5715, 16, 23, 5953, 5709, 13, 16, 51067, 23, 18024, 12, 5594, 51063, 51051, 51073, 6, 27, 15, 51071, 51072, 5983, 21, 17, 31, 51060, 18, 33, 29, 6, 36, 5936, 5705, 10, 25, 19, 50, 37, 61, 76, 75, 76, 74, 5590, 18151, 106, 96, 101, 58, 88, 53, 96, 90, 95, 50, 64, 115, 108, 65, 61, 52, 63, 32, 15, 38, 46, 95, 61, 111, 28, 68, 57, 105, 130, 133, 17, 157, 24, 38, 75, 143, 166, 67, 66, 79, 194, 139, 93, 167, 47, 71, 66, 153, 163, 46, 45, 39, 154, 43, 129, 199, 79, 173, 70, 5585, 97, 105, 72, 12, 198, 183, 174, 236, 77, 56, 73, 67, 166, 52, 26, 80, 159, 11, 220, 242, 230, 78, 119, 238, 264, 12, 184, 195, 230, 13, 15, 203, 234, 194, 89, 54, 75, 56, 185, 100, 18, 210, 180, 84, 206, 204, 210, 99, 244, 35, 164, 232, 228, 201, 60, 224, 248, 22, 41, 79, 124, 291, 299, 246, 257, 236, 111, 54, 19, 334, 193, 52, 309, 363, 194, 42, 98, 255, 247, 270, 118, 28, 252, 322, 198, 17, 231, 273, 267, 7, 237, 271, 266, 120, 117, 258, 318, 109, 91, 305, 343, 151, 55, 90, 89, 48, 92, 304, 345, 73, 302, 342, 115, 52, 286, 42, 10, 200, 199, 262, 14, 203, 263, 264, 162, 390, 225, 1024, 58, 398, 1017, 220, 134, 328, 29, 225, 332, 85, 145, 423, 56, 97, 371, 1004, 1001, 18, 264, 66, 306, 73, 96, 148, 353, 426, 98, 144, 58, 415, 109, 400, 335, 337, 172, 231, 168, 112, 176, 18176, 18164, 18183, 219, 341, 341, 437, 138, 136, 173, 337, 169, 444, 441, 222, 55, 170, 137, 1021, 192, 153, 483, 252, 494, 475, 196, 136, 507, 189, 84, 73, 135, 496, 188, 4, 54, 187, 71, 1084, 505, 258, 1119, 76, 70, 81, 38, 159, 190, 63, 24, 379, 1085, 1097, 1081, 37, 103, 388, 88, 45, 23, 158, 203, 357, 104, 1005, 152, 34, 90, 127, 352, 533, 1148, 1053, 217, 397, 46, 1002, 126, 29, 1056, 1167, 1177, 564, 100, 432, 1164, 1160, 1072, 1178, 1153, 1231, 229, 624, 470, 401, 306, 308, 387, 520, 520, 521, 432, 411, 544, 89, 419, 322, 236, 341, 5029, 50015, 538, 5021, 50024, 5005, 5046, 5024, 5028, 5013, 5026, 5054, 5028, 5036, 5038, 5063, 5043, 5046, 5049, 5053, 5042, 50054, 5094, 5065, 345, 5082, 50098, 50061, 5047, 5062, 408, 5114, 5073, 5026, 5071, 5079, 5054, 5099, 50077, 5088, 5048, 5087, 5087, 50160, 5156, 50138, 5073, 5091, 5078, 5161, 50166, 5084, 5100, 5135, 5134, 50141, 5135, 50137, 5090, 5105, 716, 5091, 5169, 5166, 748, 50181, 5131, 5245, 50249, 5111, 5103, 5161, 5123, 5160, 5240, 50218, 5269, 5259, 50262, 5180, 50265, 5320, 5159, 5215, 5199, 50273, 5374, 5167, 50271, 5385, 5159, 50363, 5392, 5168, 5391, 50389, 5188, 5283, 5198, 5295, 5268, 5218, 5407, 50405, 5186, 5301, 5298, 5222, 5302, 5308, 50435, 5454, 5260, 5449, 50444, 1056, 1200, 226, 1043, 1191, 587, 1048, 1170, 641, 1061, 1203, 224, 645, 1066, 1221, 230, 648, 60, 1049, 1227, 596, 244, 81, 588, 1232, 228, 301672111, 11, 165, 102, 1090, 6, 176, 83, 99, 1088, 431, 600, 8, 153, 105, 1086, 84, 428, 665, 1176, 1062, 283, 675, 79, 1174, 1060, 231, 637, 33, 115, 165, 301672120, 1305, 701, 56, 430, 111, 208, 1120, 163, 301672119, 1301, 719, 1051, 51, 109, 221, 57, 110, 227, 1123, 219, 656, 1302, 301672145, 1037, 1192, 713, 310, 74, 215, 1313, 717, 116, 76, 1207, 1165, 179, 52, 1107, 229, 1010, 290, 1283, 735, 1156, 1239, 234, 124, 76, 1034, 1278, 247, 777, 1013, 1345, 722, 266, 262, 1114, 74, 1226, 236, 301672139, 1365, 756, 71, 1182, 704, 267, 301672157, 664, 301672134, 381, 265, 1391, 361, 273, 751, 7007007014, 1199, 750, 301672174, 86, 156, 844, 7007007013, 7007007038, 1213, 364, 279, 850, 7007007015, 127, 112, 20, 103, 1113, 234, 97, 1121, 1191, 230, 4500833643, 102, 7, 1105, 836451, 469, 852, 301672129, 7007007020, 876, 301672127, 760, 303, 7007007052, 301672125, 775, 330, 7007007009, 7007007011, 886, 7007007048, 7007007006, 890, 7007007044, 1020, 322, 908, 7007007092, 301672124, 843, 7007007100, 1363, 7007007019, 885, 301672130, 839, 7007007105, 67, 1329, 321, 848, 340, 905, 7007007065, 7007007034, 41, 9009009018, 8008008069, 301672131, 894, 7007007069, 318, 264, 301672115, 7007007094, 396, 944, 320, 322, 323, 140, 306, 305, 949, 7007007039, 7007007181, 284, 7007007098, 1014, 7007007064, 40, 45, 36, 836453, 8008008040, 275, 30, 836479, 8008008099, 68, 5092, 5100, 5102, 5273, 5277, 5287, 5122, 5172, 5130, 5151, 5133, 5134, 5205, 5141, 5144, 5178, 5376, 5240, 5241, 5201, 5, 14, 10, 5434, 5435, 5433, 5233, 11, 14, 9, 5271, 44, 44, 45, 43, 42, 5295, 5261, 122, 107, 53, 111, 153, 157, 126, 125, 250, 5494, 77, 13, 73, 126, 42, 131, 44, 79, 135, 81, 153, 87, 89, 87, 1071, 184, 112, 115, 271, 5028, 36595, 5056, 5089, 5147, 5087, 5088, 743, 154, 1109, 926, 1174, 180, 270, 1050, 1053, 1048, 1069, 106, 8008008119, 7007007076, 7007007077, 7007007084, 5087, 20, 5229, 36592, 35360, 56, 25497, 5257, 5106, 19, 54, 5111, 5270, 5141, 5117, 5289, 5164, 5292, 5126, 122, 5092, 5303, 5038, 5009, 101, 5041, 5115, 5131, 5133, 5019, 5073, 5158, 5137, 5136, 5023, 5194, 5020, 5130, 5112, 5161, 5198, 5293, 5142, 5324, 5011, 5145, 5329, 5002, 5138, 5148, 34528, 5164, 5146, 5215, 5154, 5066, 5030, 5221, 5347, 5159, 36606, 5367, 5174, 5030, 5124, 29, 5174, 5019, 30174, 5127, 36599, 5208, 5113, 36609, 106, 36611, 5391, 5104, 5215, 169, 5213, 5137, 5140, 5185, 5088, 174, 5351, 5099, 5006, 35315, 29003, 35379, 5256, 5400, 5245, 5190, 130, 5247, 5178, 5188, 5191, 5150, 5117, 5187, 5226, 5099, 22198, 41, 5069, 5039, 5069, 5067, 6018, 5192, 5200, 5200, 31819, 5208, 5252, 5205, 5209, 5315, 5366, 5069, 5206, 5012, 5257, 5212, 176, 6044, 5401, 5213, 5419, 35367, 34543, 36784, 5098, 6033, 36785, 34540, 35365, 5171, 2080, 5256, 5266, 35338, 5267, 6067, 5437, 5330, 5221, 6092, 5035, 34525, 5105, 5225, 5237, 5449, 36613, 5045, 156, 5036, 86, 5355, 5288, 60, 41, 29886, 5243, 5213, 36593, 5047, 5244, 155, 5250, 5204, 5292, 5153, 5246, 5276, 5055, 5257, 5357, 5050, 5182, 5215, 5086, 5169, 36791, 5293, 5160, 5470, 5468, 5249, 5240, 53, 5227, 5252, 5251, 36794, 5235, 5250, 97, 36550, 56, 36603, 5299, 5305, 63, 93, 5261, 5278, 5258, 19, 5320, 5308, 11, 35329, 5032, 5498, 5312, 5497, 5496, 5315, 5500, 89, 130, 5318, 2, 6, 4, 2, 5031, 17, 9, 8, 4, 18, 22, 9, 11, 23, 9, 14, 17, 24, 29, 27, 18, 52, 39, 47, 5540, 36615, 23294, 1151, 58, 59, 5059, 9, 1240, 5558, 5259, 5530, 5346, 18, 5343, 5351, 5251, 36519, 13, 84, 82, 75, 83, 202, 76, 80, 74, 5545, 1361, 5569, 81, 90, 170, 173, 172, 173, 34541, 34712, 34539, 93, 120, 5007, 67, 35317, 32513, 34535, 5009, 35320, 239, 147, 5018, 5027, 51, 35332, 35323, 35336, 27, 87, 84, 36588, 36521, 35720, 36590, 35718, 87, 5040, 35346, 36591, 81, 5078, 36594, 5071, 124, 5040, 75, 5056, 5161, 120, 5117, 5039, 5074, 5071, 50, 5080, 19, 5129, 5190, 5197, 503, 101, 167, 5112, 5123, 5576, 5571, 5511, 1058, 1101, 5583, 1092, 57, 61, 5582, 81, 36569, 49, 98, 102, 5586, 81, 80, 85, 109, 88, 107, 104, 106, 70, 77, 79, 72, 66, 51, 108, 78, 114, 36573, 116, 114, 122, 95, 5602, 1258, 1599, 36538, 133, 132, 48, 5608, 25498, 36617, 42, 1190, 13, 137, 8, 136, 1086, 105, 3, 29, 5384, 5038, 132, 140, 5499, 141.

Distribution

US Nationwide Distribution: AL, AR, AZ, CA, CO, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NY, OK, OH, PA, PR, RI, SC, TN, TX, VA, VT, WA, WI, and WV; and the countries of: Albania, Algeria, Andorra, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, France, French Guiana, Georgia, Germany, Greece, Guadeloupe, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Maldives, Malta, Martinique, Mexico, Moldova, Monaco, Myanmar, Nepal, Netherlands, New Zealand, Norway, Oman, Pakistan, Palestine, State of, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom, Uzbekistan, Viet Nam, and Yemen.

Quantity

12,764 units