FDA Enforcement Class II Terminated

GE Healthcare, Seno Advantage 2.2 workstations with software version 22_01 and 22_02 A medical image review station that allows easy selection, processing, filming and media interchange of multi-modality images from a variety of diagnosis imaging systems.

Recall: Z-0125-2013 · Reported October 31, 2012

Enforcement

Recall Number
Z-0125-2013
Event ID
61612
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
GE Healthcare, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 31, 2012
Initiation Date
May 24, 2012
Classification Date
October 24, 2012
Termination Date
December 11, 2012
Address
3000 N Grandview Blvd, N/A, Waukesha, WI, 53188-1615, United States

Description

GE Healthcare, Seno Advantage 2.2 workstations with software version 22_01 and 22_02 A medical image review station that allows easy selection, processing, filming and media interchange of multi-modality images from a variety of diagnosis imaging systems.

Reason

GE Healthcare has recently become aware of inaccurate Scale factor annotation provided by SenoAdvantage 2.2 workstations that may impact patient safety. This issue may result in a potential miscalculation of size.

Code Info

000000010005GS 000000010004GS 000000010009GS 000000010007GS 000000010006GS 000000010010GS 000000010001GS 000000010002GS 000000010008GS 000000010003GS 00000L3A4391A7 00000L3A8D5CF0

Distribution

Worldwide Distribution -- USA (nationwide) including states of: CA, CT, NY, TX, and MS and countries of TAIWAN, ITALY, CANADA, and BAHRAIN

Quantity

12