FDA Enforcement Class II Terminated

Calix P PEEK Lumbar System, PLIF Rasp

Recall: Z-0124-2017 · Reported October 26, 2016

Enforcement

Recall Number
Z-0124-2017
Event ID
74397
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
X Spine Systems Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 26, 2016
Initiation Date
May 17, 2016
Classification Date
October 18, 2016
Termination Date
April 4, 2019
Address
452 Alexandersville Rd, N/A, Miamisburg, OH, 45342-3658, United States

Description

Calix P PEEK Lumbar System, PLIF Rasp

Reason

The trials and rasps used to prepare the surgical site for placement of the implant may become detached from the inserter assembly part and become lodged in the intervertebral space.

Code Info

Size, 22mm x 10mm x 6mm, Part #:, X034-0990, Lot #:, 015641-2-1, 015641-2-1R, 015857-24-1, 020316, 020329, 020746, 19894, UP1009, UPI002, UPI005, UPI008 & Size, 22mm x 10mm x 10mm, Part #:, X034-0994, Lot #:, 015641-3-1, 015641-3-1R,015857-25-1, 020317, 020330, 020747, UPI001, UPI004, UPI007.

Distribution

Product was shipped to the following states: AZ, CA, CO, FL, GA, ID, IN, KY, LA, MD, MI, MO, NC, NV, NY, PA, PR, TX & WV. Product was also shipped to the following countries: Australia, England, Germany, Italy, Mexico, Portugal & United Kingdom.

Quantity

182 units