Randox Lipase (Colorimetric) Product Usage: A lipase test system is a device intended to measure the activity of the enzyme lipase in serum and plasma. Lipase measurements are used in the diagnosis and treatment of diseases of the pancreas such as acute pancreatitis and obstruction of the pancreatic duct.
Enforcement
- Recall Number
- Z-0123-2017
- Event ID
- 75207
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Randox Laboratories Ltd.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- October 26, 2016
- Initiation Date
- August 2, 2016
- Classification Date
- October 18, 2016
- Termination Date
- September 7, 2017
- Address
- 55 Diamond Road, N/A, Crumlin (North), N/A, N/A, Ireland
Description
Randox Lipase (Colorimetric) Product Usage: A lipase test system is a device intended to measure the activity of the enzyme lipase in serum and plasma. Lipase measurements are used in the diagnosis and treatment of diseases of the pancreas such as acute pancreatitis and obstruction of the pancreatic duct.
According to the firm carry over was observed when the Lipase test is run directly before or after the LDL and Triglycerides assays. Customers are instructed to amend the running order for all instruments using these reagents to ensure Lipase is not place directly before or after LDL or Triglycerides. A correction was made to the IFU for the Lipase assay and all customers who purchased the Lipase kit in the USA were contacted with the updated instruction.
Catalogue number: LI3837, LI7979, LI8050, LI8361
US Nationwide Distribution