FDA Enforcement
Class III
Terminated
Alere Cholestech LDX Multianalyte Controls, PN 88769 and 88773. Enables users to monitor the performance of total cholesterol (TC), high density lipoprotein cholesterol (HDL), triglycerides (TRG) and glucose (GLU) test procedures.
Recall: Z-0122-2016
·
Reported October 21, 2015
Enforcement
- Recall Number
- Z-0122-2016
- Event ID
- 72120
- Classification
- Class III
- Status
- Terminated
- Product Type
- Devices
- Firm
- Alere San Diego, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 21, 2015
- Initiation Date
- August 27, 2015
- Classification Date
- October 15, 2015
- Termination Date
- December 22, 2015
- Address
- 9975 Summers Ridge Rd, N/A, San Diego, CA, 92121-2997, United States
Description
Alere Cholestech LDX Multianalyte Controls, PN 88769 and 88773. Enables users to monitor the performance of total cholesterol (TC), high density lipoprotein cholesterol (HDL), triglycerides (TRG) and glucose (GLU) test procedures.
Reason
Alere San Diego is recalling the Alere Cholestech LDX Multianalyte Control because the assigned control ranges for the Total Cholesterol (TC) and Triglyceride (TRG) analytes are incorrect.
Code Info
Lot C3091 and Lot C3091A
Distribution
Nationwide Distribution.
Quantity
190 kits