FDA Enforcement Class II Ongoing

P2131 Automated Field Steam Sterilizer NSN: 6530-01-641-4641

Recall: Z-0121-2022 · Reported October 27, 2021

Enforcement

Recall Number
Z-0121-2022
Event ID
88628
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Fort Defiance Industries, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
October 27, 2021
Initiation Date
August 19, 2021
Classification Date
October 15, 2021
Address
2411 Maremont Pkwy, N/A, Loudon, TN, 37774-1066, United States

Description

P2131 Automated Field Steam Sterilizer NSN: 6530-01-641-4641

Reason

The rear heater plate and cover assembly that contains the electrical wiring for the heaters does not have sufficient bonding to ground.

Code Info

Model number: P2131; NSN: 6530-01-641-4641; Serial Numbers AFS-0001 through AFS-0386. Note: AFS-0297 was not manufactured.

Distribution

Distributed nationwide to PA, MD, TX, NJ, OH, TN, UT, VA and internationally to Canada.

Quantity

380 devices