FDA Enforcement Class II Terminated

Centurion CIRCLAMP W/1.1CM BELL - single-use, disposable instrument used to compress the foreskin of the penis during circumcision of a male infant. ReOrder: 310CR

Recall: Z-0120-2020 · Reported October 16, 2019

Enforcement

Recall Number
Z-0120-2020
Event ID
83774
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Centurion Medical Products Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 16, 2019
Initiation Date
August 28, 2019
Classification Date
October 10, 2019
Termination Date
July 23, 2020
Address
100 Centurion Way, N/A, Williamston, MI, 48895-9086, United States

Description

Centurion CIRCLAMP W/1.1CM BELL - single-use, disposable instrument used to compress the foreskin of the penis during circumcision of a male infant. ReOrder: 310CR

Reason

A mismatch between the angle of the hole in the base and the angle of the bell, which may cause the clamp to remove foreskin without the use of a scalpel.

Code Info

Batch/Lot: 2019061801

Distribution

Nationwide

Quantity

1530 each