FDA Enforcement Class II Terminated

Clinical Chemistry, ABX Pentra Lipase CP, ABX Pentra 400/Pentrac400, REF: A11A01631, Model/Part # 1220001631, UDI # 03610230004944

Recall: Z-0119-2020 · Reported October 16, 2019

Enforcement

Recall Number
Z-0119-2020
Event ID
83701
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Horiba Instruments Incorporated
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 16, 2019
Initiation Date
October 30, 2018
Classification Date
October 10, 2019
Termination Date
June 16, 2020
Address
34 Bunsen, N/A, Irvine, CA, 92618-4210, United States

Description

Clinical Chemistry, ABX Pentra Lipase CP, ABX Pentra 400/Pentrac400, REF: A11A01631, Model/Part # 1220001631, UDI # 03610230004944

Reason

Potential for chemical analyzer to report abnormal results due to an interference with triglycerides CP and Cholesterol CP. An over estimation of the lipase concentration could be observed and a patient result could be declared abnormal by the instrument due to these interferences instead of normal result < 38 U/L (or < reference value used in your laboratory).

Code Info

All Lots; All Software versions

Distribution

US: AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV, WY

Quantity

1539 Chemistry Analyzers