Brand Names: Integra¿ Camino 110-4 Intracranial Pressure (ICP) Monitoring Catheters. Product / Catalogue No.s: 110-4B, 110-4BC, 110-4G, 110-4H, 110-4HC, 110-4L, 110-4LC. The fully packaged 110-4B, 110-4H, and 110-4L catheters are also provided as a component in kits that contain other cranial access devices. These kits are Catalogue No. 110-4BC, 110- 4HC, and 110-4LC respectively. Common Name: Intracranial Pressure and Pressure-Temperature Monitoring Kits Classification Name: Device, Monitoring, Intracranial Pressure 510(k) Number: K102875 The Camino 110-4 Intracranial Pressure monitoring catheters are sterile transducer-tipped pressure monitoring catheters that are used by neurosurgeons to rapidly determine and continuously monitor intracranial pressure or pressure/temperature. The Camino 110-4B catheter is used for Intracranial Pressure Monitoring in the parenchyma or subarachnoid space. The Camino 110-4G catheter is used for Post Craniotomy Subdural Pressure Monitoring. The Camino 110-4H catheter is intended for Intracranial Pressure Monitoring and Cerebrospinal fluid drainage. The Camino 110-4L catheter is intended for use with the Licox¿ Brain Oxygen Monitoring System to measure intracranial pressure in the parenchyma. The fully packaged 110-4B, 110-4H, and 110-4L catheters are also provided as a component in kits that contain other cranial access devices. These kits are Catalogue No. 110-4BC, 110- 4HC, and 110-4LC respectively.
Enforcement
- Recall Number
- Z-0117-2014
- Event ID
- 66243
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Integra Limited
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- November 6, 2013
- Initiation Date
- September 10, 2013
- Classification Date
- October 30, 2013
- Termination Date
- June 4, 2014
- Address
- 311 Enterprise Dr, N/A, Plainsboro, NJ, 08536-3344, United States
Description
Brand Names: Integra¿ Camino 110-4 Intracranial Pressure (ICP) Monitoring Catheters. Product / Catalogue No.s: 110-4B, 110-4BC, 110-4G, 110-4H, 110-4HC, 110-4L, 110-4LC. The fully packaged 110-4B, 110-4H, and 110-4L catheters are also provided as a component in kits that contain other cranial access devices. These kits are Catalogue No. 110-4BC, 110- 4HC, and 110-4LC respectively. Common Name: Intracranial Pressure and Pressure-Temperature Monitoring Kits Classification Name: Device, Monitoring, Intracranial Pressure 510(k) Number: K102875 The Camino 110-4 Intracranial Pressure monitoring catheters are sterile transducer-tipped pressure monitoring catheters that are used by neurosurgeons to rapidly determine and continuously monitor intracranial pressure or pressure/temperature. The Camino 110-4B catheter is used for Intracranial Pressure Monitoring in the parenchyma or subarachnoid space. The Camino 110-4G catheter is used for Post Craniotomy Subdural Pressure Monitoring. The Camino 110-4H catheter is intended for Intracranial Pressure Monitoring and Cerebrospinal fluid drainage. The Camino 110-4L catheter is intended for use with the Licox¿ Brain Oxygen Monitoring System to measure intracranial pressure in the parenchyma. The fully packaged 110-4B, 110-4H, and 110-4L catheters are also provided as a component in kits that contain other cranial access devices. These kits are Catalogue No. 110-4BC, 110- 4HC, and 110-4LC respectively.
Integra LifeSciences has identified through an internal evaluation that in order to comply with US FDA requirements, certain Camino¿ Intracranial Pressure Monitoring Kits require additional labelling to indicate they are UNSAFE in an MR environment. Only the following Camino¿ Intracranial Pressure Monitoring Kits are affected: 110-4B, 110-4BC, 110-4G, 110-4H, 110-4HC, 110-4L, 110-4LC.
Product / Catalogue No.s: 110-4B, 110-4BC, 110-4G, 110-4H, 110-4HC, 110-4L, 110-4LC. Lot Numbers: Catalog Number 110-4B: 3050RX195484, 3050RX198473, 305000205426, 3050RX197641, 3050RX201421, 305000205648, 305000205652, 3050RX199149, 3050RX201407, 3050RX199850, 305000205976, 3050RY189979, 3050RX200862, 305000206178, 3050RY193648, 3050RX200486, 3050RX181026, 3050RX185115, 3050RY195484, 305000207061, 305000207062, 3050RY198473, 305000207330, 305000207334, 3050RY197641, 3050RY200862, 305000207580, 305000207892, 3050RY200486, 3050RY181026, 3050RY185115, 3050RX207330, 3050RX207334, 3050RX207580, 305000208008, 305000208510, 3050RY207892, 305000210060, 3050RX208008, 3050RX208510, 3050RX205976, 3050RX206178, 305000210563, 305000211723, 3050RX210060, 305000212616, 305000213007, 305000213191, 30500X210563, 3050RX207061, 305000214590, 3050RX205426, 30500X215502, 3050RX207062, 305000215531, 30500X216410, 3050RY199850, 305000216132, 30500X216627, 30500X217119, 3050RY217112, 305000216718, 3050RX217751, 3050RX217840, 3050RY217837, 305000217858, 30500X218369, 30500X218391, 30500X219054, 3050RY206178, 305000220708, 305000218578, 30500X220430, 305000219261, 30500X221221, 305000221226, 305000219989, 30500X222569, 305000221953, 305000221955, 30500X223153, 30500X223709, 305000223275, 305000223280, 305000222640, 30500X224142, 305000223867, 305000223866, 30500X224692, 305000224352, 305000224353, 30500X225386, 305000225045, 305000225044, 30500X225992, 305000225632, 305000225633, 30500X226540, 30500X227059, 305000226844, 305000226846, 305000226127, 30500X227564, 305000226836, 30500X227820, 305000227127, 305000227128, 30500X228194, 305000227790, 305000227789, 30500X228787, 305000228522, 305000228521, 305000228971, 305000228972, 30500X229560, 305000229328, 305000229329, 30500X229766, 30500X230087, 305000229825, 305000229826, 30500X230565, 305000230303, 305000230302, 30500X231053, 305000230780, 305000230781, 30500X231786, 305000231461, 305000231462, 30500X232331, 305000232098, 305000232099, 30500X232960, 305000232605, 305000232604, 305000233245, 305000233244, 305000233243, 305000233242, 305000233241, 305000233240, 30500X233415, 30500X234183, 305000233607, 305000233608, 30500X234806, 305000234303, 305000234304, 305000235022, 305000235017, 30500X235732, 305000235598, 30500X236328, 305000236105, 305000236104, 305000236655, 305000236653, 30500Y237087, 30500X237090, 305000237232, 30500X237761, 305000237824, 305000237825, 30500X238273, 305000238263, 305000238264, 30500X239095, 305000238818, 305000238819, 30500X239497, 30500X239748, 30500X239958, 30500X240334, 30500X240668, 30500X241368, 305000241516, 305000242236, 305000242792, 305000242795, 305000242796, 305000242794, 305000242793, 305000242791, 3050RY243869, 305000243251, 305000243882, 305000243881, 305000243888, 305000243889, 305000243883, 30500X244281, 30500X244880, 305000244513,30500X245306, 305000244481, 305000244480, 305000244934, 30500X245937, 30500X246067, 305000245408, 30500X246872, 305000245835, 30500X246951, 30500X247608, 305000246460, 30500X248284, 305000247007, 3050RX227127, 3050RX234303, 305000247717, 305000249072, 3050RX242794, 3050RX242793, 3050RX237825, 3050RX241516, 30500X249488, 305000248479, 30500X250423, 305000249743, 30500X250964, 305000250456, 30500X251060, 305000251055, 30500X251560, 30500X252133, 3050RY252136, 305000251453, 30500X252771, 305000252058, 3050RX244513, 305000252646, 305000253785, 30500X254929, 3050RY255145, 30500X255141, 305000254589, 305000255549, 305000255551, 305000255038, 30500X256614, 305000256192, 305000257281, 305000256709, 305000257280, 305000258048, 30500X259385, 305000258379, 305000258921, 305000258922, 30500X259557, 305000259431, 305000259432, 30500X260159, 305000259966, 305000259960, 305000260514, 305000260527, 30500Y261748, 3050RX253785, 305000261325, 305000261326, 305000261737, 305000261736, 3050RX256192, 30500X262441, 305000262328, 305000262327, 305000262978, 305000262977, 3050RY263508, 30500X263559, 305000263504, 305000264073, 305000264074, 305000264625, 305000264624, 3050RY265348, 30500X265422, 30500X265976, 3050RY265975, 305000265530, 305000265527, 30500X266142, 305000266134, 305000266135, 30500X266789, 305000267193, 305000267194, 3050RX261325, 30500X267468, 305000267421, 305000267422, 30500X267990, 305000267673, 305000267674, 305000268294, 305000268293, 305000266784, 305000268914, 305000268915, 3050RX261736, 3050RY269955, 30500X269963, 305000269220, 305000269221, 305000270081, 3050RY271045, 30500X271047, 305000271042, 305000271716, 305000271715, 30500X272367, 305000272273, 305000272274, 305000272852, 305000272851, 30500X274204, 305000273466, 305000273463, 305000274200, 305000274201, 30500X275297, 30500X275393, 305000274816, 30500X276594, 3050RX268914, 305000277761, 3050RY282196, 305000282304, 30500X282815, 3050RY283371, 3050RX283315, 3050RX283314 Catalog Number 110-4G: 3050RX195650, 3050RX198690, 3050RY193969, 3050RY197837, 3050RX199151, 3050RX199942, 3050RX201552, 3050RY185259, 3050RX200538, 305000206177, 3050RX200989, 305000206556, 305000206557, 3050RX200990, 3050RY199151, 3050RY199942, 305000207465, 3050RY200538, 3050RY200989, 305000208007, 3050RY200990, 3050RX207465, 3050RX206557, 3050RX206556, 305000209417, 305000209105, 3050RX208007, 305000210891, 305000211419, 3050RX206177, 30500X209105, 3050RX209417, 305000212298, 30500X210891, 305000213240, 305000214103, 30500X215501, 30500X215577, 3050RX202201, 305000215628, 3050RY195650, 3050RY198690, 30500X216621, 305000216246,3050RY217113, 30500X217118, 305000217140, 30500X217835, 3050RY217839, 305000217464, 30500X218393, 305000218009, 30500X219053, 305000218225, 305000219118, 30500X220001, 30500X220429, 305000219990, 30500X221225, 305000221227, 30500X222567, 305000222642, 305000223279, 30500X223707, 305000223868, 30500X224685, 305000224354, 30500X225383, 305000225046, 305000225634, 30500X226000, 30500X226541, 305000226731, 305000226869, 305000226867, 30500X227567, 305000227129, 30500X228784, 305000228523, 305000228973, 30500X229561, 305000229330, 30500X230084, 305000229827, 305000230304, 30500X231054,305000230782, 30500X231787, 305000231463, 30500X232333, 305000232101, 30500X232776, 305000232606, 30500X233419, 305000233246, 305000233609, 30500X234690, 30500X234807, 305000234306, 305000235566, 30500X236329, 30500Y237088, 30500X237767, 305000237233, 305000238820, 30500X239496, 305000239476, 30500X240336, 30500X241357, 305000241518, 30500X242172, 305000242249, 305000243252, 305000244364, 30500X245304, 305000244935, 305000245409, 305000245838, 30500X246522, 305000246308, 30500X246873, 30500X247609, 30500X248281, 305000249073, 3050RX241518, 3050RX233246, 3050RX242249, 3050RX228973, 30500X251061, 305000251056, 30500X251563, 3050RY252137, 305000252059, 305000252648, 305000254591, 30500X254931, 30500X255143, 305000255039, 305000255548, 305000256193, 305000256737, 305000257244, 305000257246, 305000257921, 305000257920, 305000258303, 305000258290, 30500X259383, 305000258793, 305000259573, 30500X260163, 305000259958, 305000259959, 3050RX254591, 3050RX255039, 305000260513, 305000260512, 30500Y261771, 3050RX257244, 3050RX257920, 305000261159, 305000261160, 3050RY262444, 30500X262443, 305000261734, 305000261732, 305000262325, 305000262326, 3050RX258303, 305000262976, 305000262975, 3050RY263510, 30500X263560, 305000263507, 305000263506, 305000264075, 305000264076, 305000264627, 305000264628, 30500X265424, 305000265525, 305000265524, 30500X265978, 30500X266143, 305000266063, 305000266064, 30500X266787, 305000266638, 305000266637, 305000267191, 30500X267470, 305000267423, 305000267675, 3050RY269957, 30500X274206 Catalog Number 110-4H: 3050RX198281, 3050RY194144, 3050RY198281, 3050RY200861, 30500X223154, 305000220948, 305000223622, 30500X225385, 30500X233417, 305000231714, 305000241350, 30500X243874, 305000250959, 30500X254932, 305000263557 Catalog Number 110-4L: 305000221383, 30500X222566, 305000222762, 30500X223971,305000223969, 305000225078, 30500X226539, 305000226221, 30500X227058, 305000226848, 305000226749, 305000226847, 305000226871, 305000226,870, 30500X228783, 305000229450, 305000229909, 305000230871 30500X232775, 305000232199, 30500X232961, 305000233311, 305000234438, 30500X235731, 305000235181, 305000236201, 305000237368, 305000237369, 305000237799, 305000238539, 305000239102, 30500X240245, 30500X240246, 30500X240247, 305000239736, 305000239737, 30500X241365, 30500X241522, 305000241082, 30500X242170, 30500X245765, 30500X246521, 3050RX237799, 3050RX226749, 3050RX229450, 3050RX241082, 3050RX232199, 3050RY252775, 3050RX240446, 3050RY255148, 305000259553, 305000260626, 305000261835, 305000262436, 3050RX259553, 305000263146, 3050RX260626, 305000264077, 3050RX262436, 3050RX261835, 305000264175, 3050RY265351, 3050RY265766, 305000265531, 30500X266144, 3050RY266146, 30500X266522, 305000266062, 305000267297, 30500X267473, 30500X268133, 305000267784, 305000268412, 3050RX265531, 305000268916, 3050RY271046, 305000271872, 30500X272374, 305000272591, 30500X274210, 30500X274963, 305000281820 Catalog Number 110-4BC: 176044, 177018, 178013, 178898, 179757, 180139, 180955, 182279, 182908, 184079, 184687, 185284, 186277, 187293, 190452, 187780, 190451, 191891, 193414, 194074, 195161, 195608, 196987, 197492, 198189, 198512, 200069, 200490, 199609, 201039, 201595, 202423, 203565, 207077, 208532 Catalog Number 110-4HC: 179758, 180956, 183071, 184686, 186278, 187781, 190450, 192246, 194993, 195450, 198191, 199708, 201040, 201596, 202009 Catalog Number 110-4LC: 162523, 167450, 167786, 175720, 177188, 178547, 182130, 187495, 195245, 197493.
Nationwide Distribution
79,887