FDA Enforcement Class II Terminated

Elekta Medical Linear Accelerator (EMLA), configured as follows: a. Elekta Precise Digital Accelerator; b. Elekta Harmony; c. Elekta Synergy; d. Elekta Infinity; e. Versa HD

Recall: Z-0116-2022 · Reported October 20, 2021

Enforcement

Recall Number
Z-0116-2022
Event ID
88681
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Elekta, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
October 20, 2021
Initiation Date
September 17, 2021
Classification Date
October 14, 2021
Termination Date
September 30, 2024
Address
400 Perimeter Center Ter NE Ste 50, N/A, Atlanta, GA, 30346-1227, United States

Description

Elekta Medical Linear Accelerator (EMLA), configured as follows: a. Elekta Precise Digital Accelerator; b. Elekta Harmony; c. Elekta Synergy; d. Elekta Infinity; e. Versa HD

Reason

There is a risk that the Diode D1 in the Modulator can fail in a destructive manner when removed from the machine, and that the protective tube surrounding the diode may fail to contain the ejected matter.

Code Info

a. Elekta Precise Digital Accelerator - UDI GTIN Number: 5060191071017, Model Numbers: XRT 0601 (NFB), XRT 0611 (FB), All Serial Numbers; b. Elekta Harmony - UDI GTIN Number: 05060191071581, Model Numbers: XRT 2201 (NFB), All Serial Numbers; c. Elekta Synergy - UDI GTIN Number: 05060191071550, Model Numbers: XRT 0621 (NFB), XRT 0631 (FB), All Serial Numbers; d. Elekta Infinity - UDI GTIN Number: 05060191071543, 05060191071512, Model Numbers: XRT 0511 (NFB & MLCi2), XRT 0521 (FB & MLCi2), XRT 0531 (NFB & Agility), XRT 0541 (FB & Agility), All Serial Numbers; e. Versa HD - UDI GTIN Number: 05060191071574, 05060191071529, Model Numbers: XRT 2121 (NFB), XRT 2131 (FB), All Serial Numbers

Distribution

Distributed nationwide to FL, NC, OH, CA, IL, MD, OR, AL, NY, PA, MA, AK, GA, OR, KY, VA, TX, SD, AZ, NJ, MI, WI, ND, MO, WA, OK, PR, NM, NC, ID, CT, DE, LA, MT, TN, CO, UT, MN, IN, RI, IA, DC, ND, AZ, HI, KS, NE, NH, AR, WY, WV, MS, VT and internationally to Albania, Antigua/Barbuda, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegowina, Botswana, Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Croatia, Cuba, Curacao, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Finland, France, Gabon, Georgia, Germany, Ghana, Greece, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Kosovo, Kyrgystan, Lao People's Democratic Republic, Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Macedonia, Malaysia, Mexico, Morocco, Mozambique, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Pakistan, Palestinian Territory, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Republic of Serbia, Romania, Russian Federation, Rwanda, Saudi Arabia, Senegal, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Suriname, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Tunisia, Togo, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Vietnam, Yemen, Zimbabwe.

Quantity

4,405 units