Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO II TO AVVIGO INT UPGRADE US; the system utilizes an interactive display to provide physicians with physiologic, intravascular and/or intracardiac anatomical information. The system can provide various modalities which includes intravascular ultrasound (IVUS). When used with a peripheral vascular imaging catheter, the system utilizes the IVUS modality for ultrasound examination of intravascular pathology.
Enforcement
- Recall Number
- Z-0113-2025
- Event ID
- 95331
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Boston Scientific Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 23, 2024
- Initiation Date
- September 5, 2024
- Classification Date
- October 16, 2024
- Address
- 1 Scimed Pl, Maple Grove, MN, 55311-1565, United States
Description
Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO II TO AVVIGO INT UPGRADE US; the system utilizes an interactive display to provide physicians with physiologic, intravascular and/or intracardiac anatomical information. The system can provide various modalities which includes intravascular ultrasound (IVUS). When used with a peripheral vascular imaging catheter, the system utilizes the IVUS modality for ultrasound examination of intravascular pathology.
Boston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System while in either Live or Record mode. The Boston Scientific investigation determined that, due to a software anomaly, a 9-grid mark overlay is incorrectly displayed during Live mode when AVVIGO+ is connected to an OptiCross 18 Peripheral Imaging Catheter. Note that the correct overlay display for this catheter is the 15-grid mark overlay, which allows for a reference depth setting of 14.9mm for peripheral catheters. This software anomaly is limited to Live or Record mode. In Review mode, the correct 15-grid mark overlay is displayed, and the vessel image is correctly scaled to the appropriate size.
Material Number H7492493121I0, GTIN/UDI/DI 00191506033200, Batch/Serial Numbers: 102845321, 102845322, 102846834, 102846836, 102846938, 103226161, 103226162, 103226326, 103226455, 103234567, 103246416, 103246417, 103246469, 103452120, 103452131, 103452163, 103846738, 103846781, 103846785, 104508156, 104508175, 104508182, 104508183, 104520952, 104520973, 104520994, 104521036, 104534581, 104534587, 104534592, 104534595
US Nationwide distribution in Puerto Rico.