FDA Enforcement
Class II
Terminated
United Orthopedics USTAR II Knee System USTAR II Knee System Segment Part, RHS, 25 mm Length, Catalog # 2915-1025
Recall: Z-0110-2022
·
Reported October 20, 2021
Enforcement
- Recall Number
- Z-0110-2022
- Event ID
- 88692
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Uoc Usa Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 20, 2021
- Initiation Date
- August 27, 2021
- Classification Date
- October 13, 2021
- Termination Date
- January 20, 2023
- Address
- 15251 Alton Parkway Ste 100, N/A, Irvine, CA, 92618, United States
Description
United Orthopedics USTAR II Knee System USTAR II Knee System Segment Part, RHS, 25 mm Length, Catalog # 2915-1025
Reason
Due to product not meeting specification for concentricity and failing to meet requirements of dynamic fatigue testing.
Code Info
Catalog Number: 2915-1025 UDI Code: 04719872148197 Lot Code: 19E272H
Distribution
U.S. Nationwide: CA, CO, and NV No O.U.S.
Quantity
3 devices