FDA Enforcement
Class II
Ongoing
Babylog VN500 Ventilator with Installed CO2 Measurement Option-intended for the ventilation of neonatal patients from 0.4 kg (0.88 lbs) up to 10 kg (22 lbs) and pediatric patients from 5 kg (11 lbs) up to 20 kg (44 lbs) bodyweight. Catalog Number: 8417400
Recall: Z-0109-2022
·
Reported October 20, 2021
Enforcement
- Recall Number
- Z-0109-2022
- Event ID
- 88635
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Draeger Medical, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 20, 2021
- Initiation Date
- September 10, 2021
- Classification Date
- October 13, 2021
- Address
- 3135 Quarry Rd, N/A, Telford, PA, 18969-1042, United States
Description
Babylog VN500 Ventilator with Installed CO2 Measurement Option-intended for the ventilation of neonatal patients from 0.4 kg (0.88 lbs) up to 10 kg (22 lbs) and pediatric patients from 5 kg (11 lbs) up to 20 kg (44 lbs) bodyweight. Catalog Number: 8417400
Reason
Software 2.51 and Lower with Installed CO2 Measurement Option, may result in Restart of the Ventilator, brief cessation of ventilation can occur and loss of PEEP for approx. 8 seconds
Code Info
Software 2.51 and Lower
Distribution
US Nationwide
Quantity
2,169 units