FDA Enforcement Class II Ongoing

Medicina ENFit 5ml Reusable Enteral Syringe Code: LHE05

Recall: Z-0108-2023 · Reported October 26, 2022

Enforcement

Recall Number
Z-0108-2023
Event ID
90849
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Medicina Uk Ltd
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
October 26, 2022
Initiation Date
July 4, 2022
Classification Date
October 20, 2022
Address
Unit 2, Rivington View Business Park, Bolton, N/A, N/A, United Kingdom

Description

Medicina ENFit 5ml Reusable Enteral Syringe Code: LHE05

Reason

Plunger component has been manufactured with an unapproved material (Polyethylene not Polypropylene) and does not meet raw material specification

Code Info

UDI-DI: (01)05060278508153(17)250400(10)HM20055 Batch number: HM17099, HM18010, HM18052, HM18103, HM18113, HM18143, HM19009, HM19018, HM19061, HM19084, HM19132, HM19146, HM20006, HM20032, HM20055, HM20095, HM21037, HM21069, HM22019

Distribution

Nationwide

Quantity

1 box (100 syringes)