FDA Enforcement
Class II
Terminated
Carl Zeiss Meditec AG, Goeschwitzer Strasse 51-52 07745 Jena, Germany, IOLMaster 700, REF 1932-169 SN 1185393 2017-09-01
Recall: Z-0108-2019
·
Reported October 24, 2018
Enforcement
- Recall Number
- Z-0108-2019
- Event ID
- 81000
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Carl Zeiss Meditec AG
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 24, 2018
- Initiation Date
- June 25, 2018
- Classification Date
- October 15, 2018
- Termination Date
- January 15, 2021
- Address
- Carl-Zeiss-Promenade 10, N/A, Jena, N/A, N/A, Germany
Description
Carl Zeiss Meditec AG, Goeschwitzer Strasse 51-52 07745 Jena, Germany, IOLMaster 700, REF 1932-169 SN 1185393 2017-09-01
Reason
Calibration setting was incorrectly entered during manufacturing causing the optical coherence tomography (OCT) to obtain the incorrect two-dimensional images of ocular structures of the anterior and posterior segments of the eye.
Code Info
Model: 700 Catalog Number: 000000-1932-169 Serial Number: 1185393 UDI: 04049471092080
Distribution
US: CA (Mountain View and San Jose) A single system. No distributed OUS.
Quantity
1