CADD-Solis VIP ambulatory infusion pump, Model Numbers: a. 21-2120-0105-02L b. 21-2120-0105-03L c. 21-2120-0105-06L d. 21-2120-0105-07L e. 21-2120-0105-08L f. 21-2120-0105-12L g. 21-2120-0105-13L h. 21-2120-0105-14L i. 21-2120-0105-15L j. 21-2120-0105-17L k. 21-2120-0105-50L l. 21-2127-0105-02L m. 21-2127-0105-06L n. 21-2127-0105-50L
Enforcement
- Recall Number
- Z-0107-2022
- Event ID
- 88701
- Classification
- Class I
- Status
- Terminated
- Product Type
- Devices
- Firm
- Smiths Medical ASD Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- October 27, 2021
- Initiation Date
- March 24, 2020
- Classification Date
- October 19, 2021
- Termination Date
- December 22, 2023
- Address
- 6000 Nathan Ln N, N/A, Minneapolis, MN, 55442-1690, United States
Description
CADD-Solis VIP ambulatory infusion pump, Model Numbers: a. 21-2120-0105-02L b. 21-2120-0105-03L c. 21-2120-0105-06L d. 21-2120-0105-07L e. 21-2120-0105-08L f. 21-2120-0105-12L g. 21-2120-0105-13L h. 21-2120-0105-14L i. 21-2120-0105-15L j. 21-2120-0105-17L k. 21-2120-0105-50L l. 21-2127-0105-02L m. 21-2127-0105-06L n. 21-2127-0105-50L
Specific Pumps may exhibit intermittent performance in the AILD (Air in Line Detector) function. If an AILD does not sufficiently discern fluid from air in line, an air in line event may not be recognized by the pump and may not alarm to notify the clinician.
a. 21-2120-0105-02L, Serial Numbers: 1171502, 1171536, 1171537, 1171538, 1171539, 1171540, 1171541, 1171542, 1171543, 1171544, 1171545, 1171546, 1171547, 1171548, 1171549, 1171550, 1171551, 1171552, 1171553, 1171554, 1171555, 1171556, 1171557, 1171558, 1171559, 1171560, 1171561, 1171562, 1171563, 1171564, 1171565, 1171566, 1171567, 1171568, 1171569, 1171570, 1171571, 1171572, 1171573, 1171574, 1171575, 1171576, 1171577, 1171578, 1171579, 1171580, 1171581, 1171582, 1171583, 1171584, 1171632, 1171633, 1171649, 1171650, 1171665, 1171666, 1171667, 1171668, 1171669, 1171670, 1171671, 1171672, 1171673, 1171674, 1171675, 1171676, 1171677, 1171678, 1171679, 1171680, 1171681, 1171682, 1171683, 1171700, 1171701, 1171702, 1171703, 1171704, 1178267, 1178268, 1178269, 1178270, 1178271, 1178272, 1178273, 1178274, 1178275, 1178277, 1178278, 1178279, 1178280, 1178281, 1178282, 1178283, 1178284, 1178285, 1178286, 1178287, 1178288, 1178289, 1178291, 1178292, 1178293, 1178294, 1178295, 1178296, 1178297, 1178298, 1178299, 1178300, 1178301, 1178302, 1178303, 1178304, 1178305, 1178306, 1178307, 1178308, 1178309, 1178310, 1178311, 1178312, 1178313, 1178314, 1178315, 1178316, 1178317, 1178318, 1178319, 1178320, 1178321, 1178322, 1178323, 1178324, 1178325, 1178326, 1178327, 1178328, 1178329, 1178330, 1178331, 1178332, 1178333, 1178334, 1178335, 1178336, 1178337, 1178338, 1178339, 1178340, 1178341, 1178342, 1178343, 1178344, 1178345, 1178346, 1178347, 1178348, 1178349, 1178350, 1178351, 1178352, 1178353, 1178899, 1178900, 1178901, 1178902, 1178903, 1178904, 1178905, 1178907, 1178909, 1178910, 1178911, 1178912, 1178913, 1178914, 1178915, 1178916, 1178917, 1178918, 1178919, 1178920, 1178921, 1178922, 1178923, 1178924, 1178926, 1178927, 1178928, 1178929, 1178930, 1178931, 1178933, 1178934, 1178935, 1178936, 1178938, 1178939, 1178940, 1178942, 1178943, 1178945, 1178946, 1178947, 1178948, 1178949, 1178950, 1178951, 1178953, 1178956, 1178957, 1178958, 1178960, 1178961, 1178962, 1178963, 1178964, 1178965, 1178966, 1178967, 1178968, 1178969, 1178970, 1178971, 1178972, 1178973, 1178974, 1178975, 1178976, 1178977, 1178978, 1178980, 1179160, 1179162, 1179163, 1179164, 1179166, 1179167, 1179169, 1179170, 1179172, 1179173, 1179174, 1179175, 1179176, 1179178, 1179180, 1179181, 1179182, 1179185, 1179186, 1179187, 1179188, 1179189, 1179191, 1179192, 1179193, 1179195, 1179196, 1179197, 1179198, 1179199, 1179200, 1179202, 1179204, 1179205, 1179211, 1179212, 1179213, 1179214, 1179215, 1179216 b. 21-2120-0105-03L, Serial Numbers: 1171616, 1171618, 1171619, 1171620, 1171622, 1171631, 1171634, 1171635, 1171636, 1171651, 1171652, 1171653, 1171684, 1171685, 1171686, 1171687, 1171688, 1171689, 1171690, 1171691, 1171692, 1171693, 1178356, 1178357, 1178358, 1178359, 1178360, 1178361, 1178362, 1178859, 1178860, 1178862, 1178863, 1178864, 1178865, 1178866, 1178867, 1178868, 1178869, 1178870, 1178871, 1178872, 1178873, 1178874, 1178875, 1178876, 1178877, 1178878, 1178879, 1178880, 1178881, 1178882, 1178883, 1178884, 1178981, 1178982, 1178983, 1178984, 1178985, 1178986, 1178987, 1178989, 1178990, 1178991, 1178992, 1178993, 1178994, 1178995, 1178996, 1178998, 1178999, 1179000, 1179001, 1179002, 1179005, 1179006, 1179007, 1179008, 1179009, 1179010, 1179011, 1179012, 1179013, 1179015, 1179016, 1179018, 1179019, 1179020, 1179023, 1179025, 1179027, 1179028, 1179029, 1179030, 1179031, 1179032, 1179033, 1179034, 1179036, 1179037, 1179038, 1179039, 1179040, 1179041, 1179042 c. 21-2120-0105-06L, Serial Numbers: 1171405, 1171406, 1171407, 1171408, 1171409, 1171410, 1171411, 1171412, 1171413, 1171414, 1171415, 1171416, 1171417, 1171418, 1171419, 1171420, 1171421, 1171422, 1171423, 1171424, 1171425, 1171426, 1171427, 1171428, 1171429, 1171430, 1171431, 1171432, 1171433, 1171434, 1171435, 1171436, 1171437, 1171438, 1171439, 1171440, 1171441, 1171442, 1171443, 1171444, 1171445, 1171446, 1171447, 1171448, 1171449, 1171450, 1171451, 1171452, 1171453, 1171454, 1171455, 1171456, 1171457, 1171458, 1171459, 1171460, 1171461, 1171462, 1171463, 1171464, 1171465, 1171466, 1171467, 1171468, 1179142, 1179148, 1179149, 1179151, 1179152, 1179155, 1179157 d. 21-2120-0105-07L, Serial Numbers: 1171626, 1171627, 1171628, 1171629, 1171630 e. 21-2120-0105-08L, Serial Numbers: 1171469, 1171470, 1171471, 1171472, 1171473, 1171474, 1171475, 1171476, 1171477, 1171478, 1171479, 1171485, 1171486, 1171487, 1171488, 1171489, 1171490, 1171491, 1171492, 1171493, 1171494, 1171495, 1171496, 1171497, 1171498, 1171499, 1171500, 1171501, 1171503, 1171504, 1171585, 1171586, 1171587, 1171588, 1171589, 1171590, 1171591, 1171592, 1171593, 1171594, 1171598, 1171599, 1171600, 1171601, 1171602, 1171603, 1171604, 1171605, 1171606, 1171607, 1171608, 1171609, 1171610, 1171623, 1171624, 1171625, 1171643, 1171644, 1171645, 1171646, 1171647, 1171648, 1171654, 1171655, 1171656, 1171657, 1171658, 1171659, 1171660, 1171661, 1171662, 1171663, 1171664, 1171694, 1171695, 1171699 f. 21-2120-0105-12L, Serial Numbers: 1171480, 1171481 g. 21-2120-0105-13L, Serial Numbers: 1171611, 1171612, 1171613, 1171614, 1171615, 1179158 h. 21-2120-0105-14L, Serial Numbers: 1178363, 1178366, 1179096, 1179097, 1179132, 1179133, 1179135, 1179136, 1179137, 1179217, 1179242 i. 21-2120-0105-15L, Serial Numbers: 1171595, 1171596, 1171597, 1178887, 1178888, 1178889, 1178890, 1178891, 1178893, 1178894, 1178895, 1178896, 1178897, 1178898, 1179103, 1179105, 1179106, 1179107, 1179108, 1179109, 1179110, 1179112, 1179113, 1179114, 1179115, 1179116, 1179117, 1179118, 1179119, 1179120, 1179121, 1179122, 1179123, 1179124 j. 21-2120-0105-17L, Serial Numbers: 1171505, 1171506, 1171507, 1171508, 1171509, 1171510, 1171511, 1171512, 1171513, 1171514, 1171526, 1171527, 1171528, 1171529, 1171530, 1171531, 1171532, 1171533, 1171534, 1171535, 1179138, 1179139, 1179140 k. 21-2120-0105-50L, Serial Numbers: 1171516, 1171518, 1171520, 1171524, 1179098, 1179099, 1179100, 1179101, 1179102 l. 21-2127-0105-02L, Serial Numbers: 1179125, 1179127, 1179128, 1179129, 1179218, 1179219, 1179220, 1179222, 1179223, 1179224, 1179225, 1179226, 1179228, 1179230, 1179232, 1179233, 1179234, 1179235, 1179236, 1179237, 1179238, 1179239, 1179240, 1179241 m. 21-2127-0105-06L, Serial Numbers: 1171621 n. 21-2127-0105-50L, Serial Numbers: 1178354, 1178355
Distribution to European countries: BE, BH, CH, DE, DK, ES, FI, FR, GB, IT, HL, NO, PT, SE
642 units