FDA Enforcement Class II Terminated

Thoraguard Chest Tube Kit, 20 Fr

Recall: Z-0105-2022 · Reported October 20, 2021

Enforcement

Recall Number
Z-0105-2022
Event ID
88616
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Centese Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
October 20, 2021
Initiation Date
August 31, 2021
Classification Date
October 13, 2021
Termination Date
August 7, 2023
Address
4156 S 52nd St, N/A, Omaha, NE, 68117-1324, United States

Description

Thoraguard Chest Tube Kit, 20 Fr

Reason

An adhesive joint in the SmartValve component of the device can become compromised resulting in a leak.

Code Info

Model: FGS-1126 Catalog Number: TGCT120020 Lot Numbers 20100102, 21031703, and 21053102

Distribution

Distribution to a single consignee located in California.

Quantity

310 kits