FDA Enforcement Class II Ongoing

Multichem S Plus, Part Codes: a) 08P88-10, b) 08P88-11, c) 08P88-12

Recall: Z-0104-2024 · Reported October 18, 2023

Enforcement

Recall Number
Z-0104-2024
Event ID
93017
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
TECHNO-PATH MANUFACTURING LTD.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
October 18, 2023
Initiation Date
August 25, 2023
Classification Date
October 12, 2023
Address
Fort Henry Business Park, N/A, Ballina Tipperary, N/A, N/A, Ireland

Description

Multichem S Plus, Part Codes: a) 08P88-10, b) 08P88-11, c) 08P88-12

Reason

Technopath Manufacturing Ltd. has identified an issue with the glass vial commodity used to fill the Multichem S Plus Level 1, Multichem S Plus Level 2, Multichem S Plus Level 3, Multichem IA Plus and Multichem P Controls. Some customers have experienced broken vials when receiving or thawing Technopath Multichem IA Plus and Multichem S Plus Controls. Breakages have been observed upon delivery, while thawing or while opening the vial. The breakages can result in the bottom of the vial breaking off, breakage in the neck thread area or cracking on the sides of the vial. This may expose the user to biohazardous material if cut and may cause a delay in patient results.

Code Info

a) 08P88-10, UDI/DI 05391523440776, Lot Numbers: 108032101, 110072101, 119112101, 126052201, 130112201, 131112201; b) 08P88-11, UDI/DI 05391523440783, Lot Numbers: 108032102, 110072102, 119112102, 126052202, 130112202, 131112202; c) 08P88-12, UDI/DI 05391523440790, Lot Numbers: 108032103, 110072103, 119112103, 126052203, 130112203, 131112203

Distribution

US Nationwide distribution in the state of Illinois.

Quantity

117155 units