Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO + MOB SYSTEM - ZERO COST JAPAN; the system utilizes an interactive display to provide physicians with physiologic, intravascular and/or intracardiac anatomical information. The system can provide various modalities which includes intravascular ultrasound (IVUS). When used with a peripheral vascular imaging catheter, the system utilizes the IVUS modality for ultrasound examination of intravascular pathology.
Enforcement
- Recall Number
- Z-0102-2025
- Event ID
- 95331
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Boston Scientific Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 23, 2024
- Initiation Date
- September 5, 2024
- Classification Date
- October 16, 2024
- Address
- 1 Scimed Pl, Maple Grove, MN, 55311-1565, United States
Description
Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO + MOB SYSTEM - ZERO COST JAPAN; the system utilizes an interactive display to provide physicians with physiologic, intravascular and/or intracardiac anatomical information. The system can provide various modalities which includes intravascular ultrasound (IVUS). When used with a peripheral vascular imaging catheter, the system utilizes the IVUS modality for ultrasound examination of intravascular pathology.
Boston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System while in either Live or Record mode. The Boston Scientific investigation determined that, due to a software anomaly, a 9-grid mark overlay is incorrectly displayed during Live mode when AVVIGO+ is connected to an OptiCross 18 Peripheral Imaging Catheter. Note that the correct overlay display for this catheter is the 15-grid mark overlay, which allows for a reference depth setting of 14.9mm for peripheral catheters. This software anomaly is limited to Live or Record mode. In Review mode, the correct 15-grid mark overlay is displayed, and the vessel image is correctly scaled to the appropriate size.
Material Number H7492493220CZ0, GTIN/UDI/DI 00191506040482, Batch/Serial Numbers: 102991580, 102991581, 103044290, 103044291, 103044574, 103044577, 103045131, 103046258, 103049558, 103082231, 103082236, 103083046, 103138559, 103161660, 103161661, 103162801, 103163114, 103163116, 103163310, 103163312, 103426739, 103426983, 103460621, 103460622, 103542656, 103554562, 103565449, 103578369, 103648168, 103648304, 103649576, 103652219, 103652220, 103652529, 103652531, 103679467, 103679469, 103679552, 103679554, 103680003, 103683853, 103683854, 103729193, 103729194, 103732842, 103732843, 103754686, 103755483, 103755485, 103770200, 103770201, 103770291, 103770310, 103770451, 103770452, 103770453, 103770454, 103782346, 103782396, 103782397, 103782503, 103782581, 103782582, 103784279, 103798726, 103799816, 103801787, 104075451, 104075673, 104079162, 104082341, 104086595, 104086597, 104086599, 104227518, 104227521, 104228853, 104228854, 104237508, 104237509, 104237585, 104237588, 104238628, 104238629, 104238852, 104245995, 104246036, 104259376, 104423789, 104423793, 104438145, 104438152, 104972341, 104972343, 104972344, 104972345, 104975065, 104975066, 104976109, 104976110, 104977543, 104995049, 104999839, 105008687, 105008716, 105008717, 105014147, 105014722, 105014724, 105022928, 105022930, 105028203, 105028205, 105150552, 105152109, 105154989, 105161515, 105165834, 105165835, 105165836, 105165837, 105178871, 105178878, 105224149
US Nationwide distribution in Puerto Rico.
124 units