FDA Enforcement Class II Ongoing

EMPOWR 3D Knee Tibial Insert, 5R, 10MM, VE, REF: 342-10-705

Recall: Z-0102-2024 · Reported October 18, 2023

Enforcement

Recall Number
Z-0102-2024
Event ID
93098
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Encore Medical, LP
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 18, 2023
Initiation Date
August 22, 2023
Classification Date
October 12, 2023
Address
9800 Metric Blvd, N/A, Austin, TX, 78758-5445, United States

Description

EMPOWR 3D Knee Tibial Insert, 5R, 10MM, VE, REF: 342-10-705

Reason

Knee insert and acetabular system parts were swapped within inner pack prior to the product being sealed within the tray, which could result in surgeons not having the correct product for the surgery due to different anatomies (Hip vs Knee).

Code Info

UDI-DI: 00888912167321, Lot: 075T1167, Expiration: March 30, 2028

Distribution

US Nationwide distribution in the states of FL, CA, IN, HI, LA, MN, AZ, TX, NJ, PA, KS, TN, PR, SC, WA.

Quantity

36