FDA Enforcement
Class II
Terminated
cobas p 501 Post-Analytical Unit calculator/data processing module, for clinical use Product Usage: The MODULAR PRE-ANALYTICS system is a fully automated system for sample pre-processing, including centrifuging, aliquoting, and barcode labeling, prior to analysis by MODULAR SYSTEMS analyzers.
Recall: Z-0102-2017
·
Reported November 2, 2016
Enforcement
- Recall Number
- Z-0102-2017
- Event ID
- 75186
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Roche Diagnostics Operations, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- November 2, 2016
- Initiation Date
- July 15, 2016
- Classification Date
- October 25, 2016
- Termination Date
- June 5, 2017
- Address
- 9115 Hague Rd, Indianapolis, IN, 46256-1025, United States
Description
cobas p 501 Post-Analytical Unit calculator/data processing module, for clinical use Product Usage: The MODULAR PRE-ANALYTICS system is a fully automated system for sample pre-processing, including centrifuging, aliquoting, and barcode labeling, prior to analysis by MODULAR SYSTEMS analyzers.
Reason
Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning process.
Code Info
Not applicable
Distribution
US Nationwide Distribution
Quantity
9014 in total