FDA Enforcement Class II Ongoing

Access Thyroglobulin Calibrators

Recall: Z-0100-2024 · Reported October 18, 2023

Enforcement

Recall Number
Z-0100-2024
Event ID
92670
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Beckman Coulter, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
October 18, 2023
Initiation Date
July 6, 2023
Classification Date
October 11, 2023
Address
1000 Lake Hazeltine Dr, N/A, Chaska, MN, 55318-1037, United States

Description

Access Thyroglobulin Calibrators

Reason

It has been determined through an internal investigation that the alternate Access Thyroglobulin calibrator cards for use with the Thyg reagent packs distributed in April 2023 will cause a decrease of approximately 11% in patient results.

Code Info

UDI/DI 15099590227166, Serial/Lot Numbers: 233806, 234171, 234355, 234443, 234444, 234617, 337940

Distribution

worldwide except US, Canada, and China

Quantity

5980 units