FDA Enforcement Class II Terminated

Modular Pre Analytics ( EVO/Plus) Product Usage: The RSD 800A is a computer-controlled stand-online system for decapping, aliquoting and sorting of primary tubes and for producing and sorting bar coded secondary tubes.. The system operates with an online connection to the host.

Recall: Z-0098-2017 · Reported November 2, 2016

Enforcement

Recall Number
Z-0098-2017
Event ID
75186
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Roche Diagnostics Operations, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 2, 2016
Initiation Date
July 15, 2016
Classification Date
October 25, 2016
Termination Date
June 5, 2017
Address
9115 Hague Rd, Indianapolis, IN, 46256-1025, United States

Description

Modular Pre Analytics ( EVO/Plus) Product Usage: The RSD 800A is a computer-controlled stand-online system for decapping, aliquoting and sorting of primary tubes and for producing and sorting bar coded secondary tubes.. The system operates with an online connection to the host.

Reason

Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning process.

Code Info

Not applicable

Distribution

US Nationwide Distribution

Quantity

9014 in total