The GemStar Power Supply is an accessory for the GemStar Infusion Pump. It is a two piece assembled switching power supply. The Power Supply converts alternating current (AC) line input voltage of 120 or 240 V to an output to the device of 3.3 Volts DC.
Enforcement
- Recall Number
- Z-0096-2015
- Event ID
- 68047
- Classification
- Class I
- Status
- Terminated
- Product Type
- Devices
- Firm
- Hospira Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- November 5, 2014
- Initiation Date
- February 3, 2014
- Classification Date
- October 24, 2014
- Termination Date
- March 7, 2017
- Address
- 275 N Field Dr, N/A, Lake Forest, IL, 60045-2579, United States
Description
The GemStar Power Supply is an accessory for the GemStar Infusion Pump. It is a two piece assembled switching power supply. The Power Supply converts alternating current (AC) line input voltage of 120 or 240 V to an output to the device of 3.3 Volts DC.
The GemStar Infusion pump may not receive direct current (DC) power from the power supply. In one instance, smoke was observed and the GemStar pump was operating on battery power while connected to the 3 volt DC power supply.
Desktop GemStar Power Supply List Number: 13072-05; 1) US Distribution Lot Numbers: 10115G1, 11121G1, 11122G1, 1123G1, 11124G1, 12143G1, 12144G1, 12178G1, 15185G1, 12195G1, 17214G1, 18217G1, 18219G1, 19225G1, 20233G1, 21242G1, 22251G1, 23259G1, 24271G1, 25280G1, 26289G1, 26290G1, 27293G1, 27297G1, 27300G1; 2) Foreign Distribution Lot Numbers: 11128G1, 12140G1, 12142G1, 12148G1, 14176G1, 15187G1, 15196G1, 16207G1, 16208G1, 18218G1, 19226G1, 19229G1, 20234G1, 21241G1, 22255G1, 23262G1, 24270G1, 25276G1, 26285G1, 27299G1, 27301G1
Worldwide Distribution -- USA, including the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY; and, the countries of Austria, Belgium, Brazil, Columbia, France, Germany, Greece, Netherlands, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, Turkey, and United Kingdom.
US Distribution: 4,709 power supplies; Foreign Distribution: 7,425 power supplies