FDA Enforcement Class I Terminated

The GemStar Power Supply is an accessory for the GemStar Infusion Pump. It is a two piece assembled switching power supply. The Power Supply converts alternating current (AC) line input voltage of 120 or 240 V to an output to the device of 3.3 Volts DC.

Recall: Z-0095-2015 · Reported November 5, 2014

Enforcement

Recall Number
Z-0095-2015
Event ID
68047
Classification
Class I
Status
Terminated
Product Type
Devices
Firm
Hospira Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 5, 2014
Initiation Date
February 3, 2014
Classification Date
October 24, 2014
Termination Date
March 7, 2017
Address
275 N Field Dr, N/A, Lake Forest, IL, 60045-2579, United States

Description

The GemStar Power Supply is an accessory for the GemStar Infusion Pump. It is a two piece assembled switching power supply. The Power Supply converts alternating current (AC) line input voltage of 120 or 240 V to an output to the device of 3.3 Volts DC.

Reason

The GemStar Infusion pump may not receive direct current (DC) power from the power supply. In one instance, smoke was observed and the GemStar pump was operating on battery power while connected to the 3 volt DC power supply.

Code Info

Wall Mount GemStar Power Supply List Number: 13026; Lot Numbers: 25283G1, 26286G1, 26287G1, 27292G1, 27294G1, 27295G1, 27296G1

Distribution

Worldwide Distribution -- USA, including the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY; and, the countries of Austria, Belgium, Brazil, Columbia, France, Germany, Greece, Netherlands, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, Turkey, and United Kingdom.

Quantity

978 power supplies