FDA Enforcement Class II Ongoing

N Antiserum to Human IgG, 5mL Variant for BN II System and BN ProSpec System- IVD Siemens Material Number (SMN): 10446299

Recall: Z-0094-2023 · Reported October 26, 2022

Enforcement

Recall Number
Z-0094-2023
Event ID
90867
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
SIEMENS HEALTHCARE DIAGNOSTICS INC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 26, 2022
Initiation Date
August 30, 2022
Classification Date
October 14, 2022
Address
511 Benedict Ave, N/A, Tarrytown, N/A, N/A, Spain

Description

N Antiserum to Human IgG, 5mL Variant for BN II System and BN ProSpec System- IVD Siemens Material Number (SMN): 10446299

Reason

N Antiserum to Human IgG are not meeting the current High-Dose Hook Effect expectation of up to 648 mg/L for IgGU (urine) and up to 2290 mg/L for IgGC (CSF).

Code Info

UDI-DI: 0405686900179VF Lot Number: 153083; 153085; 153087A; 153090; 153095B; 153099A; 153004B. Future lots may be impacted by the issue but will then be provided with a corresponding instruction note within the reagent package

Distribution

Nationwide

Quantity

10,201 packs