FDA Enforcement
Class II
Ongoing
N Antiserum to Human IgG, 5mL Variant for BN II System and BN ProSpec System- IVD Siemens Material Number (SMN): 10446299
Recall: Z-0094-2023
·
Reported October 26, 2022
Enforcement
- Recall Number
- Z-0094-2023
- Event ID
- 90867
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- SIEMENS HEALTHCARE DIAGNOSTICS INC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 26, 2022
- Initiation Date
- August 30, 2022
- Classification Date
- October 14, 2022
- Address
- 511 Benedict Ave, N/A, Tarrytown, N/A, N/A, Spain
Description
N Antiserum to Human IgG, 5mL Variant for BN II System and BN ProSpec System- IVD Siemens Material Number (SMN): 10446299
Reason
N Antiserum to Human IgG are not meeting the current High-Dose Hook Effect expectation of up to 648 mg/L for IgGU (urine) and up to 2290 mg/L for IgGC (CSF).
Code Info
UDI-DI: 0405686900179VF Lot Number: 153083; 153085; 153087A; 153090; 153095B; 153099A; 153004B. Future lots may be impacted by the issue but will then be provided with a corresponding instruction note within the reagent package
Distribution
Nationwide
Quantity
10,201 packs