FDA Enforcement
Class II
Terminated
CVC BUNDLE VANTEX 7F, 3L, 20CM ECVC6065
Recall: Z-0094-2020
·
Reported October 16, 2019
Enforcement
- Recall Number
- Z-0094-2020
- Event ID
- 83584
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Centurion Medical Products Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 16, 2019
- Initiation Date
- April 24, 2019
- Classification Date
- October 9, 2019
- Termination Date
- April 3, 2020
- Address
- 100 Centurion Way, N/A, Williamston, MI, 48895-9086, United States
Description
CVC BUNDLE VANTEX 7F, 3L, 20CM ECVC6065
Reason
Firm distributes kits which include the BD MaxZeroNeedleess Connector, which was recalled by the manufacturer due to the potential for droplets of fluid separating from the sealed surface of the connector following disconnection.
Code Info
2018082450 2018091850 2018121850 2018122150 2019011550 2019012450 2019013050
Distribution
Nationwide domestic distribution.
Quantity
181965 total