FDA Enforcement Class II Ongoing

BD Pyxis CII Safe ES, REF: 1116-00; and BD Pyxis : CII Safe ES Desktop PC, Wired SCNR, REF: 107-254-01; CII Safe ES Desktop PC, W-LESS SCNR, BIO, REF: 107-255-01; Enterprise Server, REF: 1115-00, 134157-01, 137670-02, 134148-01, 135292-01, 136449-01, 133760-01

Recall: Z-0093-2024 · Reported October 18, 2023

Enforcement

Recall Number
Z-0093-2024
Event ID
93016
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
CareFusion 303, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 18, 2023
Initiation Date
September 6, 2023
Classification Date
October 10, 2023
Address
10020 Pacific Mesa Blvd, N/A, San Diego, CA, 92121-4386, United States

Description

BD Pyxis CII Safe ES, REF: 1116-00; and BD Pyxis : CII Safe ES Desktop PC, Wired SCNR, REF: 107-254-01; CII Safe ES Desktop PC, W-LESS SCNR, BIO, REF: 107-255-01; Enterprise Server, REF: 1115-00, 134157-01, 137670-02, 134148-01, 135292-01, 136449-01, 133760-01

Reason

When global edit is used to update multiple formulary properties simultaneously, the following properties: 1) Require Lot Number on Recall, 2) Access Destruction Bin (Witness), 3) Add to Destruction Bin (Count/Empty), 4) CII Safe Stock Out Notice are, without user warning, overwritten to the value NO , which could result in diversion of controlled substances from the automated dispensing cabinet.

Code Info

REF: 1116-00, UDI-DI: 10885403512605. Software Versions: 1.7.3 and 1.7.4 (Limited commercial release)

Distribution

US Nationwide distribution in the states of LA, NC, NY, MI, MA, NV, KS, IA, CA.

Quantity

61