FDA Enforcement Class II Ongoing

Extension Set, EXT SET W/CARESITE, 6IN., Product Code 470169, 50 units per carton. A6 Access Device is a valve intended for the aspiration, injection or gravity/pump flow of IV fluids and blood upon insertion of a male luer connector.

Recall: Z-0091-2023 · Reported October 26, 2022

Enforcement

Recall Number
Z-0091-2023
Event ID
90850
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
B. Braun Medical, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 26, 2022
Initiation Date
September 15, 2022
Classification Date
October 14, 2022
Address
901 Marcon Blvd, N/A, Allentown, PA, 18109-9512, United States

Description

Extension Set, EXT SET W/CARESITE, 6IN., Product Code 470169, 50 units per carton. A6 Access Device is a valve intended for the aspiration, injection or gravity/pump flow of IV fluids and blood upon insertion of a male luer connector.

Reason

The extension set has a high probability of failure which could lead to possible leakage or detachment at the device's bonded joint during use. The issue may result in delay of therapy, increased risk of bloodstream infections (BSI), and may expose both patients and healthcare workers to hazardous medications. Depending on the nature of the leak and type of infusion, there is also the potential for life-threatening situations to occur including air embolization or hypovolemia.

Code Info

UDI-DI (GTIN): 04046964540943 Lots/Expiration Dates 61705083 31-Oct-22 61705124 31-Oct-22 61706055 31-Oct-22 61714103 31-Dec-22 61719696 31-Jan-23 61732456 31-Mar-23 61736580 31-May-23 61739427 31-May-23 61744671 31-Jul-23 61747367 31-Aug-23 61750206 31-Aug-23 61752658 30-Sep-23 61761422 30-Nov-23 61776447 30-Apr-24 61780880 30-Apr-24 61802320 30-Sep-24 61806067 31-Oct-24 61807459 31-Oct-24 61822282 28-Feb-25

Distribution

Distribution US nationwide, Canada and Saudi Arabia.

Quantity

354,200 US; 1,800 OUS