FDA Enforcement Class II Terminated

DRX Revolution Mobile X-Ray System, Model DRXR-1, Verona Street Rochester, NY 14608 Catalog Numbers 1019397, 1023415, 1023423 --- Made in USA by Carestream Health, Inc. 150 The device is designed to perform radiographic x-ray examinations on all pediatric and adult patients, in all patient treatment areas.

Recall: Z-0090-2014 · Reported November 6, 2013

Enforcement

Recall Number
Z-0090-2014
Event ID
65805
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Carestream Health, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 6, 2013
Initiation Date
July 26, 2013
Classification Date
October 31, 2013
Termination Date
May 25, 2016
Address
150 Verona Street, N/A, Rochester, NY, 14608-1733, United States

Description

DRX Revolution Mobile X-Ray System, Model DRXR-1, Verona Street Rochester, NY 14608 Catalog Numbers 1019397, 1023415, 1023423 --- Made in USA by Carestream Health, Inc. 150 The device is designed to perform radiographic x-ray examinations on all pediatric and adult patients, in all patient treatment areas.

Reason

During an audit, Carestream discovered a DRX-1 System that captured an image which displayed an artifact. This issue was discovered in manufacturing. It occurs when the DRX product console software is updated to v5.6b software and the detector is not calibrated before use.

Code Info

Serial numbers: U.S. - 134,152,173,174,200,205,222,312,349,353,412, Worldwide Serial numbers: 469, 445, 306, 410, 453, 204, 483, 300, 171, 218

Distribution

Worldwide Distribution - US (nationwide) including the states of New York, Massachusetts, Pennsylvania, Illinois, Texas, California and Louisiana., and the countries of Australia, Canada, Finland, Italy, Switzerland and United Kingdom.

Quantity

23 units installed total (11 domestically & 12 internationally)