FDA Enforcement Class II Terminated

VITROS Chemistry Products GLU Slides, REF/Product Code 170 7801 (300 slides per sales unit), Unique Device Identifier No. 10758750009572, Rx ONLY, IVD --- For in vitro diagnostic use only. VITROS Chemistry Products GLU Slides quantitatively measure glucose (GLU) concentration in serum, plasma, urine, and cerebrospinal fluid using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System

Recall: Z-0089-2017 · Reported October 26, 2016

Enforcement

Recall Number
Z-0089-2017
Event ID
74950
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Ortho-Clinical Diagnostics
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
October 26, 2016
Initiation Date
August 11, 2016
Classification Date
October 17, 2016
Termination Date
June 11, 2018
Address
100 Indigo Creek Dr, N/A, Rochester, NY, 14626-5101, United States

Description

VITROS Chemistry Products GLU Slides, REF/Product Code 170 7801 (300 slides per sales unit), Unique Device Identifier No. 10758750009572, Rx ONLY, IVD --- For in vitro diagnostic use only. VITROS Chemistry Products GLU Slides quantitatively measure glucose (GLU) concentration in serum, plasma, urine, and cerebrospinal fluid using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System

Reason

Ortho Clinical Diagnostics confirmed that N-Acetylcysteine (NAC), a drug used for acetaminophen overdose, mucolytic therapy and is used as nutritional supplement, can interfere with VITROS GLU, LAC, TRIG, and URIC Slides assays. Ortho also tested several compounds following industry standard CLSI EP7-A2 for potential interference on VITROS GLU Slides during validation testing for a proposed change and confirmed Dextran 40 and Glutathione may interfere with current VITROS GLU Slides.

Code Info

Current/In-date GENs 10, 11, 20 through 31

Distribution

Worldwide Distribution - US (Nationwide) Puerto Rico and Bermuda), Internationally to Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Republic of Panama, Poland, Singapore, Spain, and Venezuela.

Quantity

US: 192,459 sales units; Foreign: 231,710 sales units