FDA Enforcement
Class II
Terminated
SDMS (SenTec Digital Monitoring System)
Recall: Z-0088-2017
·
Reported October 26, 2016
Enforcement
- Recall Number
- Z-0088-2017
- Event ID
- 75257
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- SenTec AG
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- October 26, 2016
- Initiation Date
- September 15, 2016
- Classification Date
- October 17, 2016
- Termination Date
- June 23, 2017
- Address
- Ringstrasse 39, N/A, Therwil, N/A, N/A, Switzerland
Description
SDMS (SenTec Digital Monitoring System)
Reason
Sentec AG found that with one batch of an internal SenTec Monitor component (Docking Station Module) there is a risk that a fastening screw could become loose under certain circumstances. A loose screw within the monitor can create uncontrolled electrical connections, including short circuits, with the risk of smolder or fire and electrical shock (patient and/or operator).
Code Info
SN 306497 to SN 306571.
Distribution
Worldwide Distribution - US Distribution to the states of : LA, MI, MO, OH, TN. and to the countries of : United Kingdom, The Netherlands and Australia.
Quantity
75 (54 US; 21 =OUS)