FDA Enforcement Class II Ongoing

Positron Emission Tomography and Computed Tomography System, Model: uMI 550

Recall: Z-0087-2023 · Reported October 19, 2022

Enforcement

Recall Number
Z-0087-2023
Event ID
90888
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Shanghai United Imaging Healthcare Co., Ltd.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 19, 2022
Initiation Date
May 30, 2022
Classification Date
October 13, 2022
Address
No. 2258 Chengbei Rd, Jiading Ind., N/A, Shanghai, N/A, N/A, China

Description

Positron Emission Tomography and Computed Tomography System, Model: uMI 550

Reason

The wireless VSM module of a mobile PET/CT system, operating in an environment with strong Wi-Fi signals, may experience ECG signal and respiratory signal loss due to Wi-Fi interference. ECG and respiratory signal loss during acquisition can result in the failure of ECG and respiratory-gated reconstruction of the PET scan, which may require rescanning of the patient.

Code Info

UDI-DI/Serial Numbers: 06971576832026/230002, 230003, 230004, 230005, 230006, 230007, 230008, 230009, 230010, 230011, 230012, 230013, 230014, 230015, 230016, 230017, 230019, 230020, 230021, 230022, 230023, 230024, 230025, 230026, 230027, 230028, 230029, 230030, 230031, 230032, 230033, 232001

Distribution

Distribution to US states of TX, WI, CA, and MI

Quantity

32