Positron Emission Tomography and Computed Tomography System, Model: uMI 550
Enforcement
- Recall Number
- Z-0087-2023
- Event ID
- 90888
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Shanghai United Imaging Healthcare Co., Ltd.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 19, 2022
- Initiation Date
- May 30, 2022
- Classification Date
- October 13, 2022
- Address
- No. 2258 Chengbei Rd, Jiading Ind., N/A, Shanghai, N/A, N/A, China
Description
Positron Emission Tomography and Computed Tomography System, Model: uMI 550
The wireless VSM module of a mobile PET/CT system, operating in an environment with strong Wi-Fi signals, may experience ECG signal and respiratory signal loss due to Wi-Fi interference. ECG and respiratory signal loss during acquisition can result in the failure of ECG and respiratory-gated reconstruction of the PET scan, which may require rescanning of the patient.
UDI-DI/Serial Numbers: 06971576832026/230002, 230003, 230004, 230005, 230006, 230007, 230008, 230009, 230010, 230011, 230012, 230013, 230014, 230015, 230016, 230017, 230019, 230020, 230021, 230022, 230023, 230024, 230025, 230026, 230027, 230028, 230029, 230030, 230031, 230032, 230033, 232001
Distribution to US states of TX, WI, CA, and MI
32