FDA Enforcement Class II Terminated

Siemens Mammomat Inspiration mammography systems. The Mammomat Inspiration system is intended for mammography exams, screening, diagnosis, and stereotactic biopsies under the supervision of medical professionals. Mammographic images can be interpreted by either hard copy film or soft copy workstation.

Recall: Z-0086-2015 · Reported October 29, 2014

Enforcement

Recall Number
Z-0086-2015
Event ID
69371
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Siemens Medical Solutions USA, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 29, 2014
Initiation Date
September 18, 2014
Classification Date
October 17, 2014
Termination Date
February 24, 2015
Address
51 Valley Stream Pkwy, Malvern, PA, 19355, United States

Description

Siemens Mammomat Inspiration mammography systems. The Mammomat Inspiration system is intended for mammography exams, screening, diagnosis, and stereotactic biopsies under the supervision of medical professionals. Mammographic images can be interpreted by either hard copy film or soft copy workstation.

Reason

The stereo biopsy devices for Mammomat Inspiration mammography systems might have integrated a safety switch which causes a failure of the functionality. The pin implemented in the safety switch may not put enough pressure on the safety circuit to prohibit movement. The needle positioning device may move even with the safety switch being set sideward. If this occurs with the needle being already i

Code Info

material #10140000, with serial numbers: 5018 3849 5042

Distribution

US Distribution in states of: NY, VA, and MA.

Quantity

3