FDA Enforcement
Class II
Ongoing
Leica Biosystems Cryostat Model Number CM1100, Material Number 14046931130, for freezing and sectioning tissue samples
Recall: Z-0085-2025
·
Reported October 23, 2024
Enforcement
- Recall Number
- Z-0085-2025
- Event ID
- 95323
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- LEICA BIOSYSTEMS NUSSLOCH GMBH
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 23, 2024
- Initiation Date
- September 4, 2024
- Classification Date
- October 15, 2024
- Address
- Heidelberger Str. 17-19, Nussloch, N/A, Germany
Description
Leica Biosystems Cryostat Model Number CM1100, Material Number 14046931130, for freezing and sectioning tissue samples
Reason
As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and therefore can cause serious injuries. The flammability is indicated on the labelling of these sprays. The cryostat labelling, however, did not include any specific warning about usage of flammable freezing sprays prior to 2019.
Code Info
UDI/DI 04049188066237, Serial Number 990
Distribution
US Nationwide distribution.
Quantity
1 unit