FDA Enforcement Class II Ongoing

Leica Biosystems Cryostat Model Number CM1100, Material Number 14046931130, for freezing and sectioning tissue samples

Recall: Z-0085-2025 · Reported October 23, 2024

Enforcement

Recall Number
Z-0085-2025
Event ID
95323
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
LEICA BIOSYSTEMS NUSSLOCH GMBH
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 23, 2024
Initiation Date
September 4, 2024
Classification Date
October 15, 2024
Address
Heidelberger Str. 17-19, Nussloch, N/A, Germany

Description

Leica Biosystems Cryostat Model Number CM1100, Material Number 14046931130, for freezing and sectioning tissue samples

Reason

As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and therefore can cause serious injuries. The flammability is indicated on the labelling of these sprays. The cryostat labelling, however, did not include any specific warning about usage of flammable freezing sprays prior to 2019.

Code Info

UDI/DI 04049188066237, Serial Number 990

Distribution

US Nationwide distribution.

Quantity

1 unit