FDA Enforcement Class II Terminated

CONCEPT HEATWAVE Electrode 130 mm x 2.3 mm diameter, REF/Catalog Number CS-023, STERILE, Rx Only Product Usage: The HeatWave Electrode is intended to be used for electro-coagulation of soft tissue in shoulder, ankle, wrist, elbow and knee arthroscopic procedures.

Recall: Z-0085-2017 · Reported October 19, 2016

Enforcement

Recall Number
Z-0085-2017
Event ID
75056
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
ConMed Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 19, 2016
Initiation Date
August 29, 2016
Classification Date
October 13, 2016
Termination Date
June 7, 2018
Address
525 French Rd, N/A, Utica, NY, 13502-5945, United States

Description

CONCEPT HEATWAVE Electrode 130 mm x 2.3 mm diameter, REF/Catalog Number CS-023, STERILE, Rx Only Product Usage: The HeatWave Electrode is intended to be used for electro-coagulation of soft tissue in shoulder, ankle, wrist, elbow and knee arthroscopic procedures.

Reason

CONMED received complaints of damage to the CS-023 electrodes shaft insulation which could result in burn marks on the shaft. If the shaft insulation is damaged in any way, it could increase the risk of patient burn.

Code Info

All lots codes manufactured from 09/01/2011 through 11/17/2015: 267221, 301001, 309889, 318652, 319649, 325019, 354759, 334627, 336041, 341704, 343927, 349952, 384294, 354768, 365239, 369773, 374480, 378521, 406137, 385695, 388538, 392382, 394178, 396542, 436808, 413702, 456341, 423877, 427704, 433503, 478513, 451923, 488101, 464062, 468012, 471297, 511327, 482366, 522335, 494840, 499981, 505321, 544240, 515060, 551152, 527734, 531802, 536700, 581072, 550988, 590037, 557461, 570449, 574167, 611804, 588892, 621304, 597982, 598682, 605515, 643396, 619686, 656909, 629202, 634911, 643395, 676005, 654991, 692617, 667118, 671471, 674410, 683144, 695062, 695063, 297982.

Distribution

Worldwide Distribution - US Nationwide in the states of: CA, HI, IL, IN, KS, LA, MA, NC, NM, NY, OH, OK, PA, TN, TX, UT, WA and WI; plus Foreign distribution to Australia, Canada, France, Germany, Italy, Spain, Czech Republic, Lithuania, Netherlands, Poland, Korea,Morocco, Malaysia, Mexico, Taiwan, and South Africa.

Quantity

Domestic: 1740; Foreign: 4770